In vivo thermoluminescence dosimetry for total body irradiation.

作者信息

Palkosková P, Hlavata H, Dvorák P, Novotný J, Novotný J

机构信息

University Hospital Motol, Department of Radiation Oncology, Prague, Czech Republic.

出版信息

Radiat Prot Dosimetry. 2002;101(1-4):597-9. doi: 10.1093/oxfordjournals.rpd.a006058.

Abstract

An improvement in the clinical results obtained using total body irradiation (TBI) with photon beams requires precise TBI treatment planning, reproducible irradiation, precise in vivo dosimetry, accurate documentation and careful evaluation. In vivo dosimetry using LiF Harshaw TLD-100 chips was used during the TBI treatments performed in our department. The results of in vivo thermoluminescence dosimetry (TLD) show that using TLD measurements and interactive adjustment of some treatment parameters based on these measurements, like monitor unit calculations, lung shielding thickness and patient positioning, it is possible to achieve high precision in absorbed dose delivery (less than 0.5%) as well as in homogeneity of irradiation (less than 6%).