Kron T, Schmiedeberg D, Schneider M, Oliver L
Institute of Radiation Oncology, Prince of Wales Hospital, Randwick, NSW.
Australas Phys Eng Sci Med. 1990 Dec;13(4):192-6.
Thermoluminescence dosimeters (TLDs) were introduced in the Radiotherapy Department of the Prince of Wales Hospital in 1975. They are used in three different types of applications: 1. to confirm doses in complicated geometries (e.g. axilla, ear, nose), 2. to measure the dose delivered to critical organs (e.g. lens, scrotum) and 3. to monitor certain treatments like total body irradiations (TBI). Using four LiF chips (TLD 100; size: 3.1mm x 3.1mm x 0.9mm) per measuring point the accuracy of each measurement is +/- 5%. Approximately 200 TLD measurements are requested from the planning radiographers per year. Lens dose determination (32% of all cases) and total body irradiations are the main applications. Since 1988 the planning radiographers have been asked to state their expectation of the dose where it could be estimated. In 14% of these cases (7% of all requests) the expected and the measured dose differed by more than 20%. This was usually followed by revising the treatment set-up and a repeated TLD measurement. For about one third of all patients more than one dose measurement is required either to monitor the treatment or to investigate changes in the delivered dose after changes in the treatment set-up. Several changes to treatment modalities (e.g. boli in various treatments, a changed scrotal shield) were made due to TLD results. This is reflected in the still increasing number of TLD requests which demonstrates the benefit of TLDs in clinical practice.
热释光剂量计(TLDs)于1975年引入威尔士亲王医院放射治疗科。它们用于三种不同类型的应用:1. 确认复杂几何形状(如腋窝、耳朵、鼻子)中的剂量;2. 测量输送到关键器官(如晶状体、阴囊)的剂量;3. 监测某些治疗,如全身照射(TBI)。每个测量点使用四个LiF芯片(TLD 100;尺寸:3.1mm×3.1mm×0.9mm),每次测量的准确度为±5%。每年规划放射技师大约会要求进行200次TLD测量。晶状体剂量测定(占所有病例的32%)和全身照射是主要应用。自1988年以来,规划放射技师被要求在可估计剂量的情况下说明他们对剂量的预期。在这些病例的14%(占所有请求的7%)中,预期剂量与测量剂量相差超过20%。这通常随后会修改治疗设置并重复进行TLD测量。对于大约三分之一的患者,需要进行不止一次剂量测量,要么是为了监测治疗,要么是为了研究治疗设置改变后输送剂量的变化。由于TLD结果,对治疗方式进行了几次改变(如在各种治疗中使用填充物、改变阴囊防护装置)。这反映在TLD请求数量仍在增加,这证明了TLD在临床实践中的益处。