Richardson Robert L, Griffiths Peter R
Department of Chemistry, University of Idaho, Moscow 83844-2343, USA.
Appl Opt. 2002 Oct 20;41(30):6332-40. doi: 10.1364/ao.41.006332.
A ray-tracing analysis of cat's-eye retroreflectors for use in active open-path Fourier-transform-infrared (OP/FT-IR) spectrometry and the results of testing f/0.5 and f/1.75 cat's-eye retroreflectors built in our laboratory with a commercial active OP/FT-IR spectrometer are presented. The ray-tracing model is based on the optical characteristics of a commercial single-telescope monostatic OP/FT-IR spectrometer and explores trends in cat's-eye behavior in practical but rigorous field conditions encountered during transportable outdoor use. All mirrors modeled are paraboloids for which the focal ratios of the primary mirror are f/0.5, f/1.75, and f/3. The effect of the focal ratio of the primary mirror, the focal length of the secondary mirror, and the off-axis alignment of the primary and the secondary mirror have been evaluated as a function of path length, including variable input-beam divergence, between the spectrometer and the cat's-eye. The paraboloidal mirrors comprising the primary and secondary of the cat's-eye retroreflectors tested were made in our laboratory by spin casting liquid epoxy-graphite composite mixtures followed by in situ polymerization with no postpolishing.
本文介绍了用于有源开路傅里叶变换红外(OP/FT-IR)光谱分析的猫眼后向反射器的光线追踪分析,以及在我们实验室中使用商用有源OP/FT-IR光谱仪对f/0.5和f/1.75猫眼后向反射器进行测试的结果。光线追踪模型基于商用单望远镜单基地OP/FT-IR光谱仪的光学特性,并探讨了在便携式户外使用过程中实际但严格的现场条件下猫眼行为的趋势。所有建模的镜子均为抛物面镜,其主镜的焦距分别为f/0.5、f/1.75和f/3。已评估了主镜焦距、副镜焦距以及主镜和副镜的离轴对准对光谱仪与猫眼之间光程长度的影响,包括可变输入光束发散。测试的猫眼后向反射器的主镜和副镜由液态环氧石墨复合混合物旋铸而成,随后进行原位聚合,无需后续抛光。