Balashov A A, Bukreev V S, Kultepin N G, Nesteruk I N, Perminov E B, Vagin V A, Zhizhin G N
Appl Opt. 1978 Jun 1;17(11):1716-22. doi: 10.1364/AO.17.001716.
The construction of a high resolution (0.005-cm(-1)) Michelson interferometer with cat's eye retroreflectors is described in some detail. The instrument line shape function of the Fourier spectrometer at 0.63 microm agrees with theory at 0.02-cm(-1) resolution. Absorption spectra of gases have been measured using a 40-m White cell. Infrared luminescence spectra have also been measured using a calcium fluoride crystal doped with divalent dysprosium, cooled to 4.2 K, and excited with radiation from a quartz halogen lamp.