Verdonschot N, Barink M, Stolk J, Gardeniers J W M, Schreurs B W
Orthopaedic Research Lab, University Medical Center Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands.
Acta Orthop Belg. 2002 Oct;68(4):348-55.
Prosthetic migration has been identified as a marker for future revision of cemented total hip reconstructions. This could be tested at a pre-clinical stage with dynamic loading experiments. The purpose of this study was to assess the effects of resting periods, which are a considerable part of the daily activity cycle, on the migration characteristics of femoral cemented stems. Ten polished Exeter stems were implanted in composite femurs and loaded either with a continuous load or a discontinuous load. Continuous loading involved 345,600 loading cycles at 1 Hz, whereas the discontinuous loading involved loading at 1 Hz for 2.5 hours and a resting period of 21.5 hours for a period of four days. Hence, a total of 36,000 loading cycles were applied to these reconstructions. The subsidence patterns of the prostheses were considerably affected by the resting periods. The prostheses exhibited a step-wise migration pattern with migration steps of about 50 microns after every resting period, whereas the continuously loaded prostheses showed a more gradual migration pattern. The final subsidence of the specimens when loaded with resting periods was significantly less than the subsidence measured without resting periods. However, these specimens were loaded with fewer loading cycles. If this was taken into account, the discontinuously loaded specimens had a four fold higher average migration per loading cycle than the specimens that were loaded continuously. In conclusion, the resting periods had a considerable effect on the migration patterns, which should be realized when these stems are analyzed with pre-clinical tests.
假体移位已被确定为骨水泥型全髋关节置换术未来翻修的一个指标。这可以在临床前阶段通过动态加载实验进行测试。本研究的目的是评估休息期(日常活动周期的重要组成部分)对股骨骨水泥柄假体移位特征的影响。将10个抛光的埃克塞特柄假体植入复合股骨中,并分别施加连续载荷或间断载荷。连续加载包括以1Hz的频率进行345,600次加载循环,而间断加载包括以1Hz的频率加载2.5小时,休息21.5小时,持续四天。因此,总共对这些假体置换施加了36,000次加载循环。休息期对假体的下沉模式有显著影响。假体呈现出阶梯式移位模式,每次休息期后移位步长约为50微米,而持续加载的假体则呈现出更渐进的移位模式。有休息期加载时标本的最终下沉明显小于无休息期时测量的下沉。然而,这些标本的加载循环次数较少。如果考虑到这一点,间断加载的标本每个加载循环的平均移位比连续加载的标本高四倍。总之,休息期对移位模式有显著影响,在通过临床前测试分析这些柄假体时应认识到这一点。