骨水泥型翻修术中Exeter柄在骨水泥内的体外循环测试。

In vitro cyclic testing of the Exeter stem after cement within cement revision.

作者信息

Wilson Lance Jon, Bell Cameron Gordon Roodveldt, Weinrauch Patrick, Crawford Ross

机构信息

Queensland University of Technology, School of Engineering Systems and Institute of Health and Biomedical Innovation, Brisbane, Australia.

出版信息

J Arthroplasty. 2009 Aug;24(5):789-94. doi: 10.1016/j.arth.2008.02.014. Epub 2008 Jun 4.

Abstract

Cement-within-cement (C-C) revision arthroplasty minimizes the complications associated with removal of secure polymethylmethacrylate. Failure at the interfacial region between new and old cement mantles remains a theoretical concern. This article assesses the cyclic fatigue properties of bilaminar cement mantles after C-C revision in vitro with the Exeter stem. Seven Exeter stems were cemented into Sawbone femurs and removed, and new undersized stems were cemented into the preserved mantle. The new constructs were loaded for 1,000,000 cycles at body temperature. Cement mantles were inspected postcycling. In no case was there delamination or failure of the cement mantle. The findings support the hypothesis that use of a thin revision cement mantle in conjunction with a polished double-tapered stem is not detrimental to the overall success of the implant. In the presence of a secure cement-bone interface in suitable patients, we recommend C-C revision techniques using double-tapered polished femoral stems.

摘要

骨水泥内骨水泥(C-C)翻修关节成形术可将与取出牢固的聚甲基丙烯酸甲酯相关的并发症降至最低。新旧骨水泥套界面区域的失效仍是一个理论上的问题。本文使用埃克塞特柄在体外评估C-C翻修术后双层骨水泥套的循环疲劳特性。将七个埃克塞特柄用骨水泥固定到Sawbone股骨中并取出,然后将新的尺寸过小的柄用骨水泥固定到保留的骨水泥套中。新构建体在体温下加载1,000,000次循环。循环后检查骨水泥套。在任何情况下,骨水泥套均未出现分层或失效。这些发现支持这样的假设,即使用薄的翻修骨水泥套结合抛光的双锥形柄对植入物的整体成功并无不利影响。在合适的患者中存在牢固的骨水泥-骨界面的情况下,我们推荐使用双锥形抛光股骨干的C-C翻修技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验