Method to correct for the effects of limited spatial resolution in phase-contrast flow MRI measurements.

作者信息

Lagerstrand K M, Lehmann H, Starck G, Vikhoff-Baaz B, Ekholm S, Forssell-Aronsson E

机构信息

Department of Radiation Physics, Sahlgrenska University Hospital, Göteborg University, Göteborg, Sweden.

出版信息

Magn Reson Med. 2002 Nov;48(5):883-9. doi: 10.1002/mrm.10288.

Abstract

Phase-contrast (PC) magnetic resonance imaging (MRI) flow measurements suffer from the effect of the point spread function (PSF) due to the limited sampling of k-space. The PSF, which in this case is a sinc function, deforms the flow profile and forms a ringing pattern around the vessel. In this work, an empirical method is presented that corrects for errors due to the deformation of the flow profile. The ringing pattern is used to obtain a well-defined vessel segmentation, which after correction provides more accurate vessel radius and volume flow rate (VFR). The correction method was developed from phantom measurements at constant flow and applied on phantom measurements at moderately pulsatile flow. After correction, the error of the estimated tube radius and the VFR was less than 10% and 5%, respectively. Corresponding errors without correction overestimated the radius by 60% and the VFR by 35%. Preliminary results indicate that the method is also valid in vivo. The variation in the estimated radius and VFR for different spatial resolution decreased when the method was applied. The presented method gives a more accurate estimation of the radius and VFR in vessels of the size of a few pixels without prior knowledge about the true vessel radius.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验