Peripheral perfusion is often poor and barely pulsatile in patients undergoing prolonged major surgery. Hence, the arterial blood oxygen saturation (SpO2) readings from commercial finger pulse oximeters can become unreliable or cease when they are most needed. To overcome this limitation, the esophagus has been investigated as an alternative measurement site, as perfusion may be preferentially preserved centrally. A reflectance esophageal pulse oximeter probe, and a processing system implemented in LabVIEW were developed. The system was evaluated in clinical measurements on 49 cardiothoracic surgery patients. The SpO2 values from the esophagus were in good agreement with arterial blood oxygen saturation (SaO2) values obtained from blood gas analysis and CO-oximetry. The means (+/-SD) of the differences between the esophageal SpO2 and SaO2 results from blood gas analysis and CO-oximetry were 0.02 +/- 0.88% and -0.73 +/- 0.72%, respectively. In five (10.2%) of the patients, the finger pulse oximeter failed for at least 10 min while the esophageal SpO2 readings remained reliable. The results confirm that the esophagus may be used as an alternative monitoring site for pulse oximetry even in patients with compromised peripheral perfusion.

摘要

在接受长时间大型手术的患者中,外周灌注通常较差且搏动微弱。因此,在最需要的时候,商用手指脉搏血氧仪测得的动脉血氧饱和度(SpO2)读数可能会变得不可靠或中断。为克服这一局限性,人们研究了将食管作为替代测量部位,因为中心部位的灌注可能会优先得到保留。开发了一种反射式食管脉搏血氧仪探头以及在LabVIEW中实现的处理系统。该系统在49例心胸外科手术患者的临床测量中进行了评估。食管测得的SpO2值与通过血气分析和一氧化碳血氧测定法获得的动脉血氧饱和度(SaO2)值高度一致。食管SpO2与血气分析和一氧化碳血氧测定法测得的SaO2结果之间差异的均值(±标准差)分别为0.02±0.88%和-0.73±0.72%。在5例(10.2%)患者中,手指脉搏血氧仪至少有10分钟无法正常工作,而食管SpO2读数仍保持可靠。结果证实,即使在外周灌注受损的患者中,食管也可作为脉搏血氧测定的替代监测部位。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验