Spurrell R A, Sowton E
J Electrocardiol. 1976;9(1):89-96. doi: 10.1016/s0022-0736(76)80017-9.
An implanted atrial synchronous pacemaker with an atrioventricular delay of 30 msec is described. This pacemaker was implanted into a patient with paroxysmal supraventricular tachycardia due to an intra AV nodal reciprocal mechanism. The pacemaker was able to trigger from atrial potentials following atrial premature beats down to a coupling time of 300 msec. Following each triggering atrial potential, a ventricular stimulus was applied 30 msec later thereby producing a ventricular premature beat in response to each sinus beat or each atrial premature beat. Retrograde conduction from this atrial premature beat blocked the re-entry mechanism within the AV node and prevented the initiation of tachycardia. A detailed discussion on all parameters of function of this pacemaker is presented.