使用Technicon H*3自动血液分析仪检测脑脊液中的白细胞计数。

Leukocyte counts in cerebrospinal fluid with the automated hematology analyzer, Technicon H*3.

作者信息

Soogarun Suphan, Sirimongkolsakul Sanong, Wiwanitkit Viroj, Siritantikorn Atchasai, Pimsane Kanokorn, Srivijarn Ritthichai

机构信息

Department of Clinical Microscopy, Faculty of Allied Health Sciences, Chulalongkorn University, Bangkok, Thailand 10330.

出版信息

Clin Lab. 2002;48(11-12):623-9.

Abstract

BACKGROUND

Presently, the counting of leukocytes in cerebrospinal fluid (CSF) is still performed manually. This requires sufficiently experienced medical technologists. At present, specific automated systems for CSF cell counting are not available. In this study, we evaluated the use of the automated hematology analyzer, Technicon H*3 in the analysis of cerebrospinal fluid (CSF).

METHODS

We tested the hematology analyzer, Technicon H3, which was not designed for CSF analysis. Sixty CSF samples were analyzed by the analyzer, Technicon H3. Then the counts obtained were compared with the manual microscopic method.

RESULTS

Linearity was tested by both plasma and normal saline solution (NSS). Concerning the NSS dilution, a good correlation can be observed at theoretical number of WBC > 30 cells/microL and so did the plasma dilution. The results of the comparison of both methods used for the determination of total leukocytes in the CSF gave a good correlation (r = 0.984). Concerning the determinations of neutrophils and lymphocytes, the correlation coefficients were 0.855 and 0.835, respectively.

CONCLUSIONS

According to our study, the use of this automated analyzer for cell counting in CSF is probably feasible. The high degree of accuracy and linearity that is offered by the analyzer should prompt us and the manufacturers to remedy the interfering factors as described by improving the algorithms. Once this is done, these analyzers may be very useful for cell counts in CSF.

摘要

背景

目前,脑脊液(CSF)中白细胞计数仍采用手工操作。这需要经验丰富的医学技术人员。目前,尚无用于脑脊液细胞计数的特定自动化系统。在本研究中,我们评估了自动血液分析仪Technicon H*3在脑脊液(CSF)分析中的应用。

方法

我们测试了并非为脑脊液分析设计的血液分析仪Technicon H*3。该分析仪对60份脑脊液样本进行了分析。然后将获得的计数结果与手工显微镜检查法进行比较。

结果

分别用血浆和生理盐水溶液(NSS)测试了线性。关于NSS稀释,当白细胞理论计数>30个/微升时可观察到良好的相关性,血浆稀释时也是如此。两种用于测定脑脊液中总白细胞的方法的比较结果具有良好的相关性(r = 0.984)。关于中性粒细胞和淋巴细胞的测定,相关系数分别为0.855和0.835。

结论

根据我们的研究,使用这种自动分析仪进行脑脊液细胞计数可能是可行的。该分析仪提供的高度准确性和线性应促使我们和制造商通过改进算法来纠正所述的干扰因素。一旦完成此操作,这些分析仪可能对脑脊液细胞计数非常有用。