Yamanishi Hachiro, Imai Nobuko, Suehisa Etsuji, Kanakura Yuzuru, Iwatani Yoshinori
Laboratory for Clinical Investigation, Osaka University Hospital, Osaka, Japan.
J Clin Lab Anal. 2007;21(5):282-5. doi: 10.1002/jcla.20184.
Manual microscopic cell counting in a Fuchs-Rosenthal (FR) chamber has been the gold standard for quantification of leukocytes in cerebrospinal fluid (CSF). However, for accurate determination of the number and differentiation of cells by chamber counting, hemocytometers must be prepared carefully and kept clean. Improper fitting of the chamber and coverslip changes the volume of sample introduced into the chamber well. Moreover, because conventional hemocytometers are used repeatedly and are breakable, there is a risk of exposure to potentially infectious material. To address these issues, disposable plastic hemocytometers have been developed. However, the accuracy, precision, and clinical usefulness of disposable chambers for CSF cells counting have not been determined. In the present study, we evaluated use of a disposable plastic counting chamber (C-Chip DHC-F01) by comparing its performance with that of an FR chamber for counting of CSF specimens and cell suspensions. Within-run precision of C-Chip counting was comparable or superior to that of FR chamber counting, and excellent correlation between cell counts obtained with the C-Chip chamber and FR chamber was observed. However, C-Chip chambers that were kept at 4 degrees C yielded significantly low cell counts. The disposable hemocytometer will reduce the risk of exposure to potentially infectious material. However, use of C-Chip chambers should be avoided in cold environments.
在富克斯-罗森塔尔(FR)计数板中进行手动显微镜细胞计数一直是脑脊液(CSF)中白细胞定量的金标准。然而,为了通过计数板计数准确确定细胞数量和分化情况,血细胞计数板必须仔细制备并保持清洁。计数板和盖玻片安装不当会改变引入计数板孔中的样本体积。此外,由于传统血细胞计数板需反复使用且易碎,存在接触潜在感染性物质的风险。为解决这些问题,已开发出一次性塑料血细胞计数板。然而,一次性计数板用于脑脊液细胞计数的准确性、精密度和临床实用性尚未确定。在本研究中,我们通过比较一次性塑料计数板(C-Chip DHC-F01)与FR计数板对脑脊液标本和细胞悬液计数的性能,评估了其使用情况。C-Chip计数的批内精密度与FR计数板相当或更优,且观察到C-Chip计数板和FR计数板获得的细胞计数之间具有良好的相关性。然而,保存在4℃的C-Chip计数板产生的细胞计数显著偏低。一次性血细胞计数板将降低接触潜在感染性物质的风险。然而,在寒冷环境中应避免使用C-Chip计数板。