Plotnikoff Nadine A, MacIntyre Donna L
School of Rehabilitation Sciences, University of British Columbia, Vancouver, Canada.
Clin J Sport Med. 2002 Nov;12(6):367-72. doi: 10.1097/00042752-200211000-00008.
To establish test-retest reliability for a strength testing protocol of the glenohumeral internal and external rotator muscles using a KinCom (Chattanooga Group, Hixson, TN) isokinetic dynamometer.
Prospective test-retest reliability study.
Rehabilitation research laboratory.
Fourteen volunteers between the ages of 27 and 35 years participated in the study. Subjects had no history of traumatic shoulder injury and no overuse-based shoulder pain interfering with activity within the last 6 months. Nine of the 14 subjects were not participating in more than 6 hours weekly of arm-based sporting activities, while the remaining five subjects were involved in swim training at workout volumes of no greater than 3000 m per week.
Subjects performed three test sessions of four maximal repetitions each of concentric and eccentric shoulder internal and external rotation bilaterally in a standardized sitting position. An average of 5.9 days occurred between consecutive test sessions (range, 2-21 days).
Average torque data were collected for concentric and eccentric internal and external rotation tests bilaterally. Intraclass correlation coefficients were calculated to compare the data across sessions. Pearson product moment correlation coefficients were calculated to assess reliability between sessions.
High test-retest reliability, ranging from 0.82 to 0.97, was demonstrated for all tests.
The results indicate that a reliable test protocol has been determined for shoulder internal and external rotation strength testing on the KinCom isokinetic dynamometer.
使用KinCom等速测力计(查塔努加集团,希克森,田纳西州)建立一种用于评估肩肱关节内、外旋转肌力量测试方案的重测信度。
前瞻性重测信度研究。
康复研究实验室。
14名年龄在27至35岁之间的志愿者参与了该研究。受试者无肩部创伤史,且在过去6个月内无因过度使用导致的肩部疼痛影响活动。14名受试者中有9人每周参与基于手臂的体育活动不超过6小时,其余5名受试者参与游泳训练,每周训练量不超过3000米。
受试者在标准坐姿下,双侧进行三次测试,每次进行四个最大重复次数的同心和偏心肩部内旋和外旋动作。连续测试之间平均间隔5.9天(范围为2至21天)。
收集双侧同心和偏心内旋和外旋测试的平均扭矩数据。计算组内相关系数以比较各测试阶段的数据。计算皮尔逊积矩相关系数以评估各测试阶段之间的信度。
所有测试均显示出较高的重测信度,范围为0.82至0.97。
结果表明,已确定了一种用于在KinCom等速测力计上进行肩部内旋和外旋力量测试的可靠测试方案。