Olsson O, Tanner K E, Ceder L, Ryd L
Department of Orthopaedics, Helsingborg Hospital, 251 87, Helsingborg, Sweden.
Int Orthop. 2002;26(6):349-55. doi: 10.1007/s00264-002-0385-x. Epub 2002 Aug 14.
The twin hook has been developed as an alternative to the conventional lag screw to be combined with a barrelled side-plate in the treatment of trochanteric hip fractures. With two oppositely directed apical hooks introduced into the subchondral bone of the femoral head, the twin hook provides different stabilising properties to the lag screw. The femoral head purchase of the twin hook and the lag screw were compared in a biomechanical study using artificial cancellous bone, and responses to axial and torsional loading was determined. A distinct yield point in load and torque was noted for the lag screw, representing failure of the laminas supporting the threads. For the twin hook, gradual increase of load and torque occurred during impaction of the bone supporting the hooks. The peak loads and torques were higher for the lag screw, but were similar for both devices after 8 mm deformation. The stiffness was higher for the lag screw, but in counter-clockwise rotation the stiffness for the lag screw was negligible. The twin hook appeared to provide fixation stability comparable to that offered by the lag screw, but with conceivable advantages in terms of a deformation response involving bone impaction and gradually increasing stability.
双钩已被开发出来,作为传统拉力螺钉的替代品,用于与筒状侧板结合治疗股骨转子间骨折。通过将两个方向相反的顶端钩插入股骨头的软骨下骨,双钩提供了与拉力螺钉不同的稳定特性。在一项使用人工松质骨的生物力学研究中,比较了双钩和拉力螺钉对股骨头的把持力,并测定了轴向和扭转载荷的响应。拉力螺钉在载荷和扭矩方面有一个明显的屈服点,代表支撑螺纹的薄片失效。对于双钩,在支撑钩的骨块嵌入过程中,载荷和扭矩逐渐增加。拉力螺钉的峰值载荷和扭矩更高,但在8毫米变形后,两种装置的峰值载荷和扭矩相似。拉力螺钉的刚度更高,但在逆时针旋转时,拉力螺钉的刚度可忽略不计。双钩似乎提供了与拉力螺钉相当的固定稳定性,但在涉及骨块嵌入和稳定性逐渐增加的变形响应方面具有可想象的优势。