Meier S J, Stockmar R
Am J Med Technol. 1976 Feb;42(2):35-8.
The bicarbonate module of the Stat/Ion was assessed in a clinical laboratory setting. Within run precision was evaluated by doing replicate analyses of patients' sera. Day-to-day precision was studied by making daily replicate analyses of buffered serum controls. The values for bicarbonate on the Stat/Ion were compared to Skeggs' method for determining carbon dioxide content on the AutoAnalyzer I. It was also compared to the value calculated from the pH and carbon dioxide pressure values obtained on the IL313-03 Blood pH/Gas system. Bilirubin interference was studied.
在临床实验室环境中对Stat/Ion的碳酸氢盐模块进行了评估。通过对患者血清进行重复分析来评估批内精密度。通过对缓冲血清对照进行每日重复分析来研究日间精密度。将Stat/Ion上碳酸氢盐的值与Skeggs在自动分析仪I上测定二氧化碳含量的方法进行比较。还将其与根据IL313 - 03血液pH/气体系统获得的pH值和二氧化碳压力值计算得出的值进行比较。研究了胆红素干扰情况。