Clinical evaluation of the Technico Stat/Ion system.

作者信息

Slaunwhite D, Clements J C, Reynoso G

出版信息

Clin Biochem. 1977 Feb;10(1):44-6. doi: 10.1016/s0009-9120(77)90638-5.

Abstract
  1. We describe our evaluation of the Technicon Stat/Ion, an instrument which performs sodium, chloride and bicarbonate analysis simultaneously. 2. All four of the assays resulted in linear response over the entire clinical range with insignificant carryover between specimens. 3. Precision studies for within-run variation were: sodium 0.3 percent, potassium 0.7 percent, chloride 0.5 percent and bicarbonate 1.6 percent. Day-to-day precision was similar to the within-run precision. 4. Comparison methods for sodium, potassium, chloride and bicarbonate utilizing flame photometry, chloridometry and titration of released carbon dioxide respectively showed the following linear regression and correlation coefficients: sodium y=0.96+5.5 (a=0.988) potassium y=1.01x+0.0 (a=.996) chloride y=0.99x+1.0 (a=.993)bicarbonate y=1.0x+1.2 (alpha=.969).
摘要
  1. 我们描述了对Technicon Stat/Ion仪器的评估,该仪器可同时进行钠、氯和碳酸氢根分析。2. 所有四项检测在整个临床范围内均产生线性响应,标本之间的残留不显著。3. 批内精密度研究结果为:钠0.3%,钾0.7%,氯0.5%,碳酸氢根1.6%。日间精密度与批内精密度相似。4. 分别采用火焰光度法、氯化物测定法和释放二氧化碳滴定法对钠、钾、氯和碳酸氢根进行比较的方法,显示出以下线性回归和相关系数:钠y = 0.96 + 5.5(a = 0.988),钾y = 1.01x + 0.0(a = 0.996),氯y = 0.99x + 1.0(a = 0.993),碳酸氢根y = 1.0x + 1.2(α = 0.969)。