Allouch C, Touzeau O, Borderie V, Puech M, Ameline B, Scheer S, Laroche L
Pôle Saint Antoine, CHNO des Quinze-Vingts, 28 rue de Charenton, 75571 Paris Cedex 12, France.
J Fr Ophtalmol. 2002 Oct;25(8):799-806.
To evaluate the accuracy and reproducibility of iridocorneal angle measurement obtained with a slit-lamp method (Orbscan).
The iridocorneal angle was measured in 100 normal eyes of 50 patients using the Orbscan II device. Angles were measured in eight meridians using both the "plane" and "polynomial" modes. For each eye, two exams were taken successively and the reproducibility was calculated by analyzing the difference between both measurements. Angle value was correlated with subjective spherical equivalent and other biometric measurements obtained by Orbscan (i.e., anterior chamber depth, axial length, corneal diameter).
Reproducibility of the angle measurement was significantly better with the plane mode than with the polynomial mode (5.4%+/-4.7 versus 7.2%+/-6.2; p<0.001). Reproducibility significantly varied with localization (p<0.001). Significant differences were found between reconstruction modes (49.3 degrees +/-4.6 degrees for plane mode versus 42.8 degrees +/-6.7 degrees for the polynomial mode; p<0.001) and according to localization (p=0.02). Correlation with the other biometric measurements was higher for the plane mode than for the polynomial mode. Angle measurement with the plane mode showed significant correlation with subjective spherical equivalent (r(s)=-0.59; p<0.001), the anterior chamber depth measurement (r(s)=-0.74; p<0.001), the axial length measurement (r(s)=-0.34; p<0.002), and the corneal diameter measurement (r(s)=-0.34; p<0.002).
The iridocorneal angle measure obtained with the plane mode showed high reproducibility and it significantly correlated with clinical parameters. It appears to be useful for clinical application.
评估使用裂隙灯方法(Orbscan)测量虹膜角膜角的准确性和可重复性。
使用Orbscan II设备对50例患者的100只正常眼睛进行虹膜角膜角测量。使用“平面”和“多项式”模式在八个子午线上测量角度。对每只眼睛连续进行两次检查,并通过分析两次测量之间的差异来计算可重复性。角度值与主观球镜当量以及通过Orbscan获得的其他生物测量值(即前房深度、眼轴长度、角膜直径)相关。
平面模式下角度测量的可重复性明显优于多项式模式(5.4%±4.7对7.2%±6.2;p<0.001)。可重复性随定位显著变化(p<0.001)。在重建模式之间(平面模式为49.3度±4.6度,多项式模式为42.8度±6.7度;p<0.001)以及根据定位(p=0.02)发现了显著差异。平面模式与其他生物测量值的相关性高于多项式模式。平面模式下的角度测量与主观球镜当量(r(s)=-0.59;p<0.001)、前房深度测量(r(s)=-0.74;p<0.001)、眼轴长度测量(r(s)=-0.34;p<0.002)和角膜直径测量(r(s)=-0.34;p<0.002)显示出显著相关性。
平面模式下获得的虹膜角膜角测量显示出高可重复性,并且与临床参数显著相关。它似乎对临床应用有用。