de Mattos Ivanildo Luiz, Gorton Lo, Ruzgas Tautgirdas
Department of Analytical Chemistry, University of Lund, P.O. Box 124, S-22100 Lund, Sweden.
Biosens Bioelectron. 2003 Mar;18(2-3):193-200. doi: 10.1016/s0956-5663(02)00185-9.
Gold (Au) and platinum (Pt) screen-printed electrodes were modified with Prussian Blue (PB) for the development of amperometric sensors selective for hydrogen peroxide detection. The sensors exhibited sensitivities towards H(2)O(2) equal to 2 A M(-1) cm(-2) for Au and 1 A M(-1) cm(-2) for Pt electrodes. The sensors were also employed as the basis for construction of glucose biosensors through further modification with crystallised glucose oxidase immobilised in a Nafion membrane. In order to improve the operational stability of the modified electrodes a buffer solution containing tetrabutylammonium toluene-4-sulfonate was used. The long-term performance of the sensors and biosensors were evaluated by continuous monitoring of hydrogen peroxide and glucose solutions (50 microM and 1 mM, respectively) in the flow-injection mode for 10 h.
金(Au)和铂(Pt)丝网印刷电极用普鲁士蓝(PB)进行修饰,以开发用于选择性检测过氧化氢的电流型传感器。这些传感器对过氧化氢的灵敏度,金电极等于2 A M⁻¹ cm⁻²,铂电极等于1 A M⁻¹ cm⁻²。通过用固定在Nafion膜中的结晶葡萄糖氧化酶进一步修饰,这些传感器还被用作构建葡萄糖生物传感器的基础。为了提高修饰电极的操作稳定性,使用了含有四丁基甲苯-4-磺酸盐的缓冲溶液。通过在流动注射模式下连续监测过氧化氢和葡萄糖溶液(分别为50 μM和1 mM)10小时,评估了传感器和生物传感器的长期性能。