Golara Moneli, Jones Clare, Randhawa Manjit, Shennan Andrew H
Maternal and Foetal Research Unit, GKT School of Medicine, St Thomas' Hospital, London, UK.
Blood Press Monit. 2002 Dec;7(6):325-8. doi: 10.1097/00126097-200212000-00006.
To evaluate the accuracy of the OMRON-MIT, (an automated inflationary oscillometric blood pressure monitoring device) for use in the general population by direct comparison to standard auscultatory mercury sphygmomanometry.
Two trained, blinded observers taking simultaneous blood pressure measurements alternating between a mercury sphygmomanometer and the OMRON-MIT.
The evaluation was carried out according to the revised British Hypertension Society protocol. The validation data was also classified by the American Association for the Advancement of Medical Instrumentation (AAMI).
The OMRON-MIT device achieved a BHS grade A for both systolic blood pressure (SBP) and diastolic blood pressure (DBP). It also met the criteria for the AAMI protocol, the mean difference between the MIT and standard was -2.0 +/- 7 mmHg for SBP and -2.0 +/- 6 mmHg for DBP.
Inflationary oscillometry is a valid technique for the measurement of blood pressure. The OMRON-MIT, which uses this technique, is accurate and can be recommended for clinical use in the adult population.
通过与标准听诊式汞柱血压计直接比较,评估欧姆龙MIT(一种自动充气振荡式血压监测设备)在普通人群中的准确性。
两名经过培训的、不知情的观察者同时进行血压测量,在汞柱血压计和欧姆龙MIT之间交替使用。
根据修订后的英国高血压学会方案进行评估。验证数据也由美国医学仪器促进协会(AAMI)进行分类。
欧姆龙MIT设备在收缩压(SBP)和舒张压(DBP)方面均达到了英国高血压学会A级标准。它也符合AAMI方案的标准,MIT与标准之间的平均差值,收缩压为-2.0±7 mmHg,舒张压为-2.0±6 mmHg。
充气振荡法是一种有效的血压测量技术。采用该技术的欧姆龙MIT准确可靠,可推荐用于成年人群的临床使用。