Haus A G, Metz C E, Chiles J T, Rossmann K
Radiology. 1976 Mar;118(3):705-9. doi: 10.1148/118.3.705.
The measured x-ray spectra from a molybdenum anode Senographe mammography unit and a tungsten anode unit were used to calculate the x-ray energy spectra transmitted through various thicknesses of fat and water (breast-equivalent materials). The dependence of subject contrast and patient exposure on (a) x-ray spectra, (b) attenuation properties of two breast-equivalent materials, and (c) thickness of breast-equivalent material was predicted. Radiographs of resected breast tissue confirmed these predictions and demonstrated the general relative effects of x-ray beam quality on image contrast in mammography.