钼靶管和钨靶管产生的X射线光谱对乳腺摄影图像质量的影响。

The effect of x-ray spectra from molybdenum and tungsten target tubes on image quality in mammography.

作者信息

Haus A G, Metz C E, Chiles J T, Rossmann K

出版信息

Radiology. 1976 Mar;118(3):705-9. doi: 10.1148/118.3.705.

Abstract

The measured x-ray spectra from a molybdenum anode Senographe mammography unit and a tungsten anode unit were used to calculate the x-ray energy spectra transmitted through various thicknesses of fat and water (breast-equivalent materials). The dependence of subject contrast and patient exposure on (a) x-ray spectra, (b) attenuation properties of two breast-equivalent materials, and (c) thickness of breast-equivalent material was predicted. Radiographs of resected breast tissue confirmed these predictions and demonstrated the general relative effects of x-ray beam quality on image contrast in mammography.

摘要

利用钼靶乳腺摄影设备的钼阳极和钨阳极设备测得的X射线光谱,计算透过不同厚度脂肪和水(乳腺等效材料)的X射线能谱。预测了受检者对比度和患者照射量对(a)X射线光谱、(b)两种乳腺等效材料的衰减特性以及(c)乳腺等效材料厚度的依赖性。切除乳腺组织的X光片证实了这些预测,并展示了X射线束质量对乳腺摄影图像对比度的一般相对影响。