The comparative performance of mammographic X-ray systems that use different anode/filter combinations has been assessed for screen-film and digital imaging. Monte Carlo techniques have been used to calculate average glandular dose as well as contrast and signal-to-noise ratio for imaging two test details. Five anode/filter combinations have been studied to establish the potential for dose saving or image quality improvement. For screen-film mammography, it was found that little benefit is gained by changing from a standard 28 kV molybdenum/molybdenum spectrum for breasts up to 6 cm thick. For thicker breasts, where the tube potential for the standard technique might be increased, 20% improvement in contrast can be achieved without dose penalty using molybdenum/rhodium or rhodium/rhodium spectra, whereas dose savings of more than 50% can be attained whilst maintaining contrast using tungsten/rhodium or rhodium/aluminium spectra. In digital mammography, a molybdenum/molybdenum spectrum delivers the lowest dose for a 2 cm breast, but gives the highest dose for thicker breasts. Tungsten/rhodium or rhodium/aluminium spectra provide the lowest doses at greater thicknesses. It is concluded that for screen-film mammography, molybdenum/molybdenum is the spectrum of choice for all but the thickest or most glandular breasts. In digital mammography, an alternative spectrum is preferable for breasts thicker than 2 cm.

摘要

对于使用不同阳极/滤过组合的乳腺X射线摄影系统,已针对屏-片成像和数字成像评估了其比较性能。已使用蒙特卡罗技术来计算平均腺体剂量以及对两个测试细节进行成像时的对比度和信噪比。已研究了五种阳极/滤过组合,以确定节省剂量或改善图像质量的潜力。对于屏-片乳腺摄影,发现对于厚度达6 cm的乳房,从标准的28 kV钼/钼能谱转换几乎没有益处。对于更厚的乳房,标准技术的管电压可能会提高,使用钼/铑或铑/铑能谱可在不增加剂量的情况下使对比度提高20%,而使用钨/铑或铑/铝能谱在保持对比度的同时可节省超过50%的剂量。在数字乳腺摄影中,钼/钼能谱对于2 cm厚的乳房剂量最低,但对于更厚的乳房剂量最高。钨/铑或铑/铝能谱在更大厚度时提供最低剂量。结论是,对于屏-片乳腺摄影,除了最厚或腺体最多的乳房外,钼/钼能谱是首选。在数字乳腺摄影中,对于厚度超过2 cm的乳房,选择其他能谱更合适。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验