Quantitative analysis of clindamycin in human plasma by liquid chromatography/electrospray ionisation tandem mass spectrometry using d1-N-ethylclindamycin as internal standard.

作者信息

Rechberger Gerald N, Fauler Günter, Windischhofer Werner, Köfeler Harald, Erwa Wolfgang, Leis Hans-Jörg

机构信息

University Children's Hospital, Division of Analytical Biochemistry and Mass Spectrometry, Auenbruggerplatz 30, 8036 Graz, Austria.

出版信息

Rapid Commun Mass Spectrom. 2003;17(2):135-9. doi: 10.1002/rcm.887.

Abstract

A new method for the quantitative analysis of clindamycin in human plasma by liquid chromatography/electrospray ionisation tandem mass spectrometry (LC/ESI-MS/MS) is presented. Recently published methods possess a disadvantage because of their use of internal standards with extraction and ionisation properties different from those of clindamycin. To avoid these problems, d(1)-N-ethylclindamycin was synthesised for use as internal standard by N-demethylation and subsequent d(1)-N-ethylation. Plasma sample preparation was done by an easy and rapid liquid-liquid extraction using ethyl acetate. The method was validated in the expected concentration range for a pharmacokinetic study. Calibration graphs were linear within the range 0.05-3.2 microg/mL plasma. Intra-day precision was between 0.90% (2.8 microg/mL) and 3.25% (0.05 microg/mL), inter-day variability was found to be between 1.33% (0.7 microg/mL) and 2.60% (0.05 microg/mL). Inter-day accuracy showed deviations between 0.4% (0.05 microg/mL) and -4.8% (0.2 microg/mL). The method is simple and robust, and has been applied to the batch analysis of clindamycin during a pharmacokinetic study.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验