Abstract

The minimum voltage needed to perform laparoscopic sterilization was studied to learn more about prevention of bowel burns. Using a "CUT" waveform at very low levels, it is practical to coagulate fallopian tubes with as little as 100 volts peak. If electrosurgical cutting of fulguration (charring) is desired, the nimimum voltage is about 700 volts peak. The changes in electrical resistance of tissue were measured during coagulation of sample tissues and actual sterilizations. A discussion of basic electrosurgical principles is included.

摘要

为了更多地了解预防肠道烧伤,对进行腹腔镜绝育所需的最低电压进行了研究。使用极低水平的“切割”波形,仅用100伏峰值电压就可以实际凝固输卵管。如果需要进行电外科切割或烧灼(碳化),最低电压约为700伏峰值。在对样本组织进行凝固和实际绝育过程中测量了组织电阻的变化。文中还包括了对基本电外科原理的讨论。