[Comparison of temperature measurements during laparoscopic sterilization and the histologic findings of the cauterized segments of the fallopian tube (author's transl)].

作者信息

Larbig J, Göltner E, Bässler R

出版信息

Geburtshilfe Frauenheilkd. 1975 Mar;35(3):190-3.

PMID:124285
Abstract

During laparotomies the temperature generated by tubal sterilization with the laparoscopic equipment was measured continually in 32 fallopian tubes. The thermal damage to the fallopian tube caused by different energies and coagulation durations was examined microscopically. During the coagulation of the fallopian tube, with an energy of 20-30 watt temperatures of around 100 degrees C are produced at the point of coagulation. The effective coagulation time is dependant upon the energy created and the duration of the coagulation. The best histologic results were obtained with a duration of the coagulation of 60 seconds at an energy of 30 watts. This corresponds to an effective coagulation time of 50 seconds. The temperature of the coagulation instrument becomes normal within 1 to 2 minutes after the coagulation. If the electric energy is increased to 50 or 100 watt, sparking is observed almost consistently. With these energies, temperatures up to 600 degrees C are observed. The temperature of the metal parts of the coagulation forceps then become normal after 3-4 minutes. In order to avoid burns during laparoscopic tubal sterilizations and in order to obtain a good histologic result the procedure should be done with an electrical energy of 30 watt and a duration of the coagulation of 60 seconds.

摘要

相似文献

引用本文的文献

1
[The kidney after thermonecrosis. Cell-kinetical investigations on rats (author's transl)].
Virchows Arch A Pathol Anat Histol. 1977 Oct 7;375(4):319-29. doi: 10.1007/BF00427062.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验