使用三种自动分析系统对13项临床实验室检测结果进行比较。

Comparison of results for 13 clinical laboratory determinations with three automated analytical systems.

作者信息

Sampson E J, Derck D D, Demers L M

出版信息

Clin Chem. 1976 Mar;22(3):346-9.

PMID:1253409
Abstract

We evaluated the Abbott Bichromatic Analyzer-100 (ABA-100) for use in the routine clinical chemistry laboratory by examining 13 different determinations that can be performed on the instrument. Results with the Du Pont "aca" and Technicon continuous-flow systems were compared to the ABA-100 in terms of upper limits of linearity, inter-run coefficient of variation, and results for samples from patients. The upper limits of linearity for the methods on the ABA-100 exceeded all of those for the continuous-flow systems, except for urea nitrogen. Precision of the ABA-100 was as good as or better than that of the aca for all determinations, except for glucose in a normal control serum and creatine kinase and creatinine in an above-normal control serum.

摘要

我们通过检测可在雅培双色分析仪 - 100(ABA - 100)上进行的13种不同检测项目,对其用于常规临床化学实验室进行了评估。将杜邦“aca”和Technicon连续流动系统的检测结果在检测线性上限、批间变异系数以及患者样本检测结果方面与ABA - 100进行了比较。ABA - 100上各检测方法的线性上限,除尿素氮外,均超过了连续流动系统的所有线性上限。除正常对照血清中的葡萄糖以及高于正常对照血清中的肌酸激酶和肌酐外,ABA - 100在所有检测项目中的精密度与aca相当或优于aca。