Xu S, Chen R, Chen F, Hu Z, Ding M, Lu S
Department of Cardiovascular Surgery, 2nd Affiliated Hospital of Zhejiang Medical University, Hangzhou 310009.
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 1999 Jun;16(2):240-2.
This study sought to address the question, "will the phenomenon of 'total erythrocyte destruction' come about during a long time of in vitro hemolysis test by roller pump". One-day-old boving blood was employed in the test by Polystan pump(group A) and Cobe pump (group B). TMB method was used to measure the level of plasma-free hemoglobin. The samples of plasma-free hemoglobin were taken at the time before pumping and pumping for 4,6,8,10, 12,14 and 16 hours. The index of hemolysis was calculated according to Koller's formula. The results showed that the levels of plasma-free hemoglobin in the two groups both increased linearly during the puming of blood. The index of hemolysis was 0.2960 mg/L group A, and 0.3993 mg/L in group B; no statistically significant difference was noted between the two groups. These data indicated that no phenomenon of "total erythrocyte destruction" was observed in this hemolysis test by roller pump.
本研究旨在探讨“在长时间体外溶血试验中,滚压泵是否会导致‘全红细胞破坏’现象”这一问题。采用Polystan泵(A组)和Cobe泵(B组)对一日龄牛血进行试验。采用TMB法测定血浆游离血红蛋白水平。在泵血前以及泵血4、6、8、10、12、14和16小时采集血浆游离血红蛋白样本。根据科勒公式计算溶血指数。结果显示,两组在泵血过程中血浆游离血红蛋白水平均呈线性升高。A组溶血指数为0.2960 mg/L,B组为0.3993 mg/L;两组间差异无统计学意义。这些数据表明,在本次滚压泵溶血试验中未观察到“全红细胞破坏”现象。