Joos P E, Godoi A F L, De Jong R, de Zeeuw J, Van Grieken R
Micro- and Trace Analysis Center, Department of Chemistry, University of Antwerp, Universiteitsplein 1. 2610 Antwerp, Belgium.
J Chromatogr A. 2003 Jan 24;985(1-2):191-6. doi: 10.1016/s0021-9673(02)01843-5.
A rapid determination of benzene, toluene, ethylbenzene and the three xylene isomers (BTEX), including a nearly baseline separation of the xylene isomers in environmental samples within 1 min has been carried out using low-pressure gas chromatography-ion trap mass spectrometry (LP-GC-IT-MS). In order to evaluate the different parameters which may influence the performance of LP-GC-IT-MS, different column and mass spectral parameters were varied. Comparing LP-GC-IT-MS with the conventional equivalent, we obtained excellent detection limits as well as a good RSD of 8-13% in ition to a much shorter analysis time. In order to evaluate LP-GC-IT-MS for use in environmental samples, we determined BTEX in air.
使用低压气相色谱-离子阱质谱法(LP-GC-IT-MS)对苯、甲苯、乙苯和三种二甲苯异构体(BTEX)进行了快速测定,包括在1分钟内对环境样品中的二甲苯异构体进行近乎基线的分离。为了评估可能影响LP-GC-IT-MS性能的不同参数,对不同的色谱柱和质谱参数进行了变化。将LP-GC-IT-MS与传统等效方法进行比较,我们获得了出色的检测限,以及8-13%的良好相对标准偏差(RSD),此外分析时间也短得多。为了评估LP-GC-IT-MS在环境样品中的应用,我们测定了空气中的BTEX。