Analytical Chemistry Department, University of Valencia, Edifici Jeroni Muñoz, 50 Dr. Moliner, 46100 Burjassot, Valencia, Spain.
J Chromatogr A. 2009 Nov 27;1216(48):8549-56. doi: 10.1016/j.chroma.2009.10.001. Epub 2009 Oct 9.
A new procedure for the passive sampling in air of benzene, toluene, ethylbenzene and xylene isomers (BTEX) is proposed. A low-density polyethylene layflat tube filled with a mixture of solid phases provided a high versatility tool for the sampling of volatile compounds from air. Several solid phases were assayed in order to increase the BTEX absorption in the sampler and a mixture of florisil and activated carbon provided the best results. Direct head-space-gas chromatography-mass spectrometry (HS-GC-MS) measurement of the whole deployed sampler was employed for a fast determination of BTEX. Absorption isotherms were used to develop simple mathematical models for the estimation of BTEX time-weighted average concentrations in air. The proposed samplers were used to determine BTEX in indoor air environments and results were compared with those found using two reference methodologies: triolein-containing semipermeable membrane devices (SPMDs) and diffusive Radiello samplers. In short, the developed sampling system and analytical strategy provides a versatile, easy and rapid atmospheric monitor (VERAM).
提出了一种用于空气中苯、甲苯、乙苯和二甲苯异构体(BTEX)被动采样的新方法。填充有混合固相的低密度聚乙烯平纹管为从空气中采样挥发性化合物提供了一种高通用性工具。为了增加采样器中 BTEX 的吸收,对几种固相进行了测定,结果表明,弗罗里硅土和活性炭的混合物效果最佳。采用直接顶空-气相色谱-质谱法(HS-GC-MS)对整个展开的采样器进行直接测量,可快速测定 BTEX。采用吸附等温线法建立了简单的数学模型,用于估算空气中 BTEX 的时间加权平均浓度。所提出的采样器用于测定室内空气环境中的 BTEX,结果与两种参考方法(含三油酸甘油酯的半渗透膜装置(SPMDs)和扩散式 Radiello 采样器)发现的结果进行了比较。总之,所开发的采样系统和分析策略提供了一种通用、简便、快速的大气监测器(VERAM)。