Rahmatalla Muhib F, Zuberbuhler Peter C, Lange Carlos F, Finlay Warren H
Aerosol Research Laboratory of Alberta, Department of Mechanical Engineering, Edmonton, Canada.
J Aerosol Med. 2002 Winter;15(4):379-85. doi: 10.1089/08942680260473452.
Experimental work has been conducted on the effect of an add-on holding chamber (Aerochamber) on the characteristics of deposition in a mouth-throat model using 100-microg hydrofluoroalkane-beclomethazone dipropionate (QVAR) metered dose inhalers at inhalation flow rates of 28.3, 60, and 90 L/min. A filter or cascade impactor downstream of the mouth-throat collected aerosol not depositing. The results emphasize the important well documented effect of a valved holding chamber (VHC), in reducing drug deposition in the mouth-throat. This reduction is largest (24% of nominal dose) at the lowest flow rate tested, becoming insignificant at 60 L/min. Total amount of drug delivered distal to the mouth-throat increases with flow rate both with and without a holding chamber, increasing from 42% to 69% of the nominal dose without a VHC as the inspiratory flow rate increases from 28.3 to 90 L/min. The effect of the holding chamber on post mouth-throat delivery was small, reaching significance only at the highest flow rate (90 L/min), where an increase by 8% of the nominal dose was observed. No significant effect on MMAD of beclomethasone-dipropionate occurred when the holding chamber was used. An argument based on the interaction between induced turbulence and particle inertia is used to shed light on the above observations.
已经开展了实验工作,研究了附加储雾罐(爱全乐储雾罐)对使用100微克氢氟烷烃-二丙酸倍氯米松(信必可都保)定量吸入器在口咽模型中沉积特性的影响,吸入流速分别为28.3、60和90升/分钟。口咽下游的过滤器或级联冲击器收集未沉积的气溶胶。结果强调了带阀储雾罐(VHC)在减少口咽药物沉积方面已得到充分记录的重要作用。在测试的最低流速下,这种减少幅度最大(占标称剂量的24%),在60升/分钟时变得不显著。无论有无储雾罐,输送到口咽远端的药物总量均随流速增加,在没有VHC的情况下,随着吸气流速从28.3升/分钟增加到90升/分钟,输送到口咽远端的药物总量从标称剂量的42%增加到69%。储雾罐对口咽后输送的影响较小,仅在最高流速(90升/分钟)时达到显著水平,此时观察到标称剂量增加了8%。使用储雾罐时,对二丙酸倍氯米松的质量中值空气动力学直径没有显著影响。基于诱导湍流和颗粒惯性之间的相互作用进行了论证,以阐明上述观察结果。