Ferretti A, Conteduca F, Morelli F, Monteleone L, Nanni F, Valente M
Department of Orthopaedic Surgery, Hospital S. Andrea, University La Sapienza, Via Lidia, 73, 00179, Rome, Italy.
Int Orthop. 2003;27(1):22-5. doi: 10.1007/s00264-002-0395-8. Epub 2002 Aug 13.
We studied the biomechanical properties of a twisted doubled semitendinosus and gracilis graft. We applied an un-axial load in order to reproduce the kinematics of a reconstructed anterior cruciate ligament (ACL). A modified cryo-jaw clamp system was used to minimize soft tissue slippage. The lower grip, after fixation of the free ends of the tendons, was rotated 45 degrees, translated 1 cm, and bent 40 degrees, simulating a knee sprain. The graft was tested to failure using a servohydraulic machine. The specimen from one knee of seven unembalmed cadavers was assigned to the untwisted (parallel) bundles group, while its pair was assigned to the twisted group. For the parallel bundles group, the mean maximum load was 1,709.3+/-581.9 N, for the twisted group 2,428.3+/-475.4 N (P<0.05). The mean stiffness was respectively 213.6+/-72.4 N/mm and 310.3+/-97.3 N/mm (P=0.08). Although caution should be used in extrapolating the results to clinical estimates of the strength of hamstring grafts, the results of the present study could justify the use of twisted semitendinosus and gracilis bundles in ACL reconstruction.
我们研究了双股半腱肌与股薄肌扭转肌腱移植体的生物力学特性。为了重现重建前交叉韧带(ACL)的运动学,我们施加了单轴载荷。使用改良的低温夹钳系统来尽量减少软组织滑动。在固定肌腱游离端后,下部夹具旋转45度、平移1厘米并弯曲40度,模拟膝关节扭伤。使用伺服液压机对移植体进行直至破坏的测试。从7具未防腐尸体的一侧膝关节获取的标本被分配到未扭转(平行)束组,而其配对标本被分配到扭转组。平行束组的平均最大载荷为1709.3±581.9牛,扭转组为2428.3±475.4牛(P<0.05)。平均刚度分别为213.6±72.4牛/毫米和310.3±97.3牛/毫米(P=0.08)。尽管在将结果外推至腘绳肌腱移植体强度的临床评估时应谨慎,但本研究结果可为在ACL重建中使用扭转的半腱肌和股薄肌束提供依据。