Urlando Alessandro, Dempster Philip, Aitkens Susan
Life Measurement Incorporated, Concord, CA 94520, USA.
Pediatr Res. 2003 Mar;53(3):486-92. doi: 10.1203/01.PDR.0000049669.74793.E3.
ABSTRACT This article introduces and evaluates the PEA POD Infant Body Composition System, an air displacement plethysmograph designed for the assessment of body composition in infants between birth and 6 mo of age. The performance of the PEA POD was evaluated by repeated testing of National Institute of Standards and Technology-traceable weights and volumes. Mass was measured in a single session. Volume was measured in four sessions during a 2-d period (five times/session for both). The mean values for repeated mass measurements were almost identical to the masses of traceable weights. The SD and CV for repeated volume measurements were 1.1-4.5 mL and 0.02-0.09%, respectively. Both the mean SD and CV were within very narrow ranges (1.4-3.1 mL for SD and 0.03-0.08% for CV) across all volume levels. Furthermore, mean CV values using results from the four sessions indicated excellent within- and between-day reliability. Regression analyses (by session or with all sessions combined) of the measured volume against actual volume gave very low standard error of the estimate (SEE) (0.853-1.957 mL) and very high R(2) (1.000), with the intercept and slope not significantly different from 0 and 1, respectively. The mean percentage error in volume measurements was < or =0.05% at all volume levels. The study findings and the operational and physical characteristics of the system indicate that the PEA POD has the potential to provide clinicians and researchers with a diagnostic and research tool that is accurate, easily used by operators, and comfortable for subjects.
摘要 本文介绍并评估了PEA POD婴儿身体成分测量系统,这是一种用于评估出生至6个月龄婴儿身体成分的空气置换体积描记仪。通过对可溯源至美国国家标准与技术研究院的重量和体积进行重复测试,对PEA POD的性能进行了评估。在单次测量中测定质量。在2天内分4次测量体积(每次测量5次)。重复测量质量的平均值与可溯源重量的质量几乎相同。重复测量体积的标准差(SD)和变异系数(CV)分别为1.1 - 4.5 mL和0.02 - 0.09%。在所有体积水平下,平均SD和CV均在非常窄的范围内(SD为1.4 - 3.1 mL,CV为0.03 - 0.08%)。此外,使用4次测量结果计算的平均CV值表明,日内和日间可靠性极佳。对测量体积与实际体积进行回归分析(按测量次数或合并所有测量次数),得到的估计标准误差(SEE)非常低(0.853 - 1.957 mL),决定系数R²非常高(1.000),截距和斜率分别与0和1无显著差异。在所有体积水平下,体积测量的平均百分比误差≤0.05%。研究结果以及该系统的操作和物理特性表明,PEA POD有潜力为临床医生和研究人员提供一种准确、易于操作人员使用且受试者感觉舒适的诊断和研究工具。