Reliability and validity of the lung volume measurement made by the BOD POD body composition system.

作者信息

Davis James A, Dorado Silvia, Keays Kathleen A, Reigel Kimberly A, Valencia Kristoffer S, Pham Patrick H

机构信息

Department of Kinesiology, Laboratory of Applied Physiology, California State University/Long Beach, Long Beach, CA 90840-4901, USA.

出版信息

Clin Physiol Funct Imaging. 2007 Jan;27(1):42-6. doi: 10.1111/j.1475-097X.2007.00713.x.

Abstract

The BOD POD Body Composition System uses air-displacement plethysmography to measure body volume. To correct the body volume measurement for the subject's lung volume, the BOD POD utilizes pulmonary plethysmography to measure functional residual capacity (FRC) at mid-exhalation as that is the subject's lung volume during the body volume measurement. Normally, FRC is measured at end-exhalation. The BOD POD FRC measurement can be corrected to an end-exhalation volume by subtracting approximately one-half of the measured tidal volume. Our purpose was to determine the reliability and validity of the BOD POD FRC measurement at end-exhalation. Ninety-two healthy adults (half female) underwent duplicate FRC measurements by the BOD POD and one FRC measurement by a traditional gas dilution technique. The latter method was used as the reference method for the validity component of the study. The order of the FRC measurements by the two methods was randomized. The test-retest correlation coefficients for the duplicate BOD POD FRC measurements for the male and female subjects were 0.966 and 0.948, respectively. The mean differences between the BOD POD FRC trial #1 measurement and gas dilution FRC measurement for the male and female subjects were -32 and -23 ml, respectively. Neither difference was statistically significant. The correlation coefficients for these two measurements in the male and female subjects were 0.925 and 0.917, respectively. Based on these results, we conclude that the BOD POD FRC measurement in healthy males and females is both reliable and valid.

摘要

BOD POD人体成分分析系统采用空气置换体积描记法来测量身体体积。为了校正受测者的肺体积对身体体积测量结果的影响,BOD POD利用肺体积描记法在呼气中期测量功能残气量(FRC),因为这是身体体积测量期间受测者的肺体积。通常,FRC是在呼气末测量的。通过减去大约一半测得的潮气量,可将BOD POD的FRC测量值校正为呼气末体积。我们的目的是确定BOD POD在呼气末测量FRC的可靠性和有效性。92名健康成年人(一半为女性)接受了BOD POD对FRC的重复测量,以及通过传统气体稀释技术进行的一次FRC测量。后一种方法用作该研究有效性部分的参考方法。两种方法对FRC的测量顺序是随机的。男性和女性受测者BOD POD对FRC重复测量的重测相关系数分别为0.966和0.948。男性和女性受测者BOD POD FRC试验1测量值与气体稀释FRC测量值之间的平均差异分别为-32和-23毫升。两者差异均无统计学意义。男性和女性受测者这两项测量的相关系数分别为0.925和0.917。基于这些结果,我们得出结论,BOD POD对健康男性和女性FRC的测量既可靠又有效。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验