使用区域生物电阻抗分析测量小腿电阻率的方法及可重复性

Methods and reproducibility of measurement of resistivity in the calf using regional bioimpedance analysis.

作者信息

Zhu Fansan, Sarkar Shubho, Kaitwatcharachai Charoen, Greenwood Roger, Ronco Claudio, Levin Nathan W

机构信息

Renal Research Institute and Division of Nephrology and Hypertension, Beth Israel Medical Center, New York, NY, USA.

出版信息

Blood Purif. 2003;21(1):131-6. doi: 10.1159/000067855.

Abstract

UNLABELLED

The usefulness of regional bioimpedance analysis (RBIA) in determining the dry weight in dialysis patients is currently being investigated. The aim of this study was to evaluate the reproducibility of measurement of resistivity in the calf.

METHODS

Twenty-five normal subjects and 10 patients undergoing regular hemodialysis were studied. Four electrodes inserted into a blood pressure cuff were placed on the calf. Bioimpedance was continuously measured over 3- to 5-min periods (Xitron Hydra). After a resting period of 1 min, cuff pressure was increased to above the systolic blood pressure (SBP) for a few seconds to expel excess ECF fluid and then deflated. The effect on recordings of moving the electrodes 2 cm higher and then 2 cm lower than the normal position was studied in 8 normal subjects. In a second study of reproducibility, post-dialysis measurements were made twice in 10 patients who maintained the same post-dialysis target weight throughout the study period.

RESULTS

The mean resting resistivity rho(0) in normal subjects was 532.6 +/- 95 omega.cm rising to 583.9 +/- 99.7 omega.cm when cuff pressure was applied rho(p). The average values of rho(0) and rho(p) in patients post-dialysis were 489 +/- 74 and 537 +/- 77 respectively showing that there were no significant differences in rho(0) and rho(p) between normal subjects and patients post-dialysis. The mean values of change in resistivity when the electrodes were shifted between the lowest and highest positions on the calf were -3.66 +/- 4.45 and -1.44 +/- 3.82%, respectively. Repeat measurement of resistivity in patients post-dialysis varied by 2.04 +/- 2.29% while post-dialysis body weight varied by 0.17 +/- 0.47%.

CONCLUSION

In this study, resistivity measurement by RBIA at the calf showed similar levels of fluid loading in patients post-dialysis as in normal subjects. This study also showed that change in electrode position resulted in a mean change in resistivity of <5% and repeated measurements showed a change in resistivity <3% while body weight changes were <0.2%. This technique appears to have an acceptable level of reproducibility for its application to the assessment of patient hydration.

摘要

未标注

目前正在研究区域生物电阻抗分析(RBIA)在确定透析患者干体重方面的实用性。本研究的目的是评估小腿部电阻测量的可重复性。

方法

对25名正常受试者和10名接受定期血液透析的患者进行了研究。将插入血压袖带的四个电极置于小腿上。在3至5分钟内连续测量生物电阻抗(Xitron Hydra)。在休息1分钟后,将袖带压力升高至收缩压(SBP)以上几秒钟以排出多余的细胞外液(ECF),然后放气。在8名正常受试者中研究了将电极从正常位置向上移动2厘米然后向下移动2厘米对记录的影响。在第二项可重复性研究中,对10名在整个研究期间维持相同透析后目标体重的患者进行了两次透析后测量。

结果

正常受试者的平均静息电阻ρ(0)为532.6±95Ω·cm,施加袖带压力ρ(p)时升至583.9±99.7Ω·cm。透析后患者的ρ(0)和ρ(p)平均值分别为489±74和537±77,表明正常受试者和透析后患者的ρ(0)和ρ(p)无显著差异。当电极在小腿最低和最高位置之间移动时,电阻变化的平均值分别为-3.66±4.45%和-1.44±3.82%。透析后患者电阻的重复测量变化为2.04±2.29%,而透析后体重变化为0.17±0.47%。

结论

在本研究中,通过RBIA测量小腿部电阻显示,透析后患者的液体负荷水平与正常受试者相似。本研究还表明,电极位置的改变导致电阻平均变化<5%,重复测量显示电阻变化<3%,而体重变化<0.2%。这项技术在应用于评估患者水合状态时似乎具有可接受的可重复性水平。