Suppr超能文献

使用胰岛素类似物持续皮下胰岛素输注治疗的1型糖尿病患者中动态血糖监测系统的临床性能

Clinical performance of CGMS in type 1 diabetic patients treated by continuous subcutaneous insulin infusion using insulin analogs.

作者信息

Guerci Bruno, Floriot Michèle, Böhme Philip, Durain Danielle, Benichou Muriel, Jellimann Stéphanie, Drouin Pierre

机构信息

Service de Diabétologie, Maladies Métaboliques et Maladies de la Nutrition, Centre d'Investigation Clinique-INSERM/CHU de NANCY-Hôpital Jeanne d'Arc, Centre Hospitalo-Universitaire de Nancy, Toul cedex, France.

出版信息

Diabetes Care. 2003 Mar;26(3):582-9. doi: 10.2337/diacare.26.3.582.

Abstract

OBJECTIVE

Performance criteria have been established for in vitro blood glucose monitoring, particularly for the self-monitoring of blood glucose using glucose meters. Devices intended for use in the future, such as the continuous glucose monitoring system (CGMS), should satisfy similar criteria, particularly in diabetic patients under intensive therapy.

RESEARCH DESIGN AND METHODS

The analysis was conducted on 18 type 1 diabetic patients (not controlled, HbA(1c) >7.5%) treated by external pump using insulin analogs. Each patient received a glucose sensor for 3 days during his/her hospitalization and was instructed in its operation. Medtronic criteria were used to determine the accuracy of the CGMS. In addition, the data were analyzed according to American Diabetes Association (ADA) criteria, Clarke Error Grid analysis, and method of residuals, with the glucose oxidase method using a Beckman analyzer used as the reference method. Specificity and sensitivity were evaluated from the viewpoint of accuracy in the detection of hypoglycemia. For nine patients, two glucose sensors were simultaneously inserted into an abdominal site to determine the reproducibility of the system. RESULTS-Among the 33 glucose sensors inserted, 6 (18%) were nonfunctional. The mean duration of CGMS recording was 63 +/- 12 h. From all of the 692 sets of data that paired glucose readings and CGMS, the coefficients of correlation ranged from 0.87 to 0.92 and the mean absolute error ranged from 12.8 to 15.7%. The time experienced in hypoglycemia (<55 mg/dl) was reported at 86 +/- 62 min/day. Only 39% of the CGMS values satisfied the ADA precision criteria to within +/-10%, and 19% of these values satisfied the future ADA precision criteria of accuracy to within +/-5%. The means of difference method showed that the CGMS slightly underestimated the plasma glucose values (mean = -12 mg/dl). Error grid analysis showed only 77% of the glucose sensor values were in zone A, and 98.9% were in zones A and B. Two values fell in zone C and a single value fell in zone D. The sensitivity and specificity of the CGMS to detect hypoglycemia were 33 and 96%, respectively. A total of 6666 paired sensor values were recorded with a coefficient of correlation of 0.84 with a coefficient of variation of 8.25%.

CONCLUSIONS

CGMS could be useful in routine clinical practice to provide much more information on the glucose profile than intermittent self-monitoring of blood glucose (SMBG). However, CGMS cannot be used as a replacement for glucose meters because it does not satisfy the conventional performance goals set down for in vitro glucose measurements and could therefore lead to clinically incorrect treatment decisions.

摘要

目的

已制定了体外血糖监测的性能标准,特别是使用血糖仪进行血糖自我监测的标准。未来打算使用的设备,如连续血糖监测系统(CGMS),应满足类似标准,尤其是在强化治疗的糖尿病患者中。

研究设计与方法

对18例使用胰岛素类似物进行体外泵治疗的1型糖尿病患者(未得到控制,糖化血红蛋白>7.5%)进行分析。每位患者在住院期间接受了3天的葡萄糖传感器,并接受了操作指导。采用美敦力标准来确定CGMS的准确性。此外,根据美国糖尿病协会(ADA)标准、克拉克误差网格分析和残差法对数据进行分析,以使用贝克曼分析仪的葡萄糖氧化酶法作为参考方法。从低血糖检测准确性的角度评估特异性和敏感性。对9例患者,将两个葡萄糖传感器同时插入腹部部位以确定系统的可重复性。结果——在插入的33个葡萄糖传感器中,6个(18%)无法正常工作。CGMS记录的平均持续时间为63±12小时。从所有配对血糖读数和CGMS的692组数据来看,相关系数范围为0.87至0.92,平均绝对误差范围为12.8%至15.7%。报告的低血糖(<55mg/dl)时间为86±62分钟/天。只有39%的CGMS值符合ADA精度标准,误差在±10%以内,其中19%的值符合未来ADA精度标准中的±5%以内准确性要求。差异均值法显示CGMS略微低估了血浆葡萄糖值(均值=-12mg/dl)。误差网格分析显示只有77% 的葡萄糖传感器值处于A区,98.9% 处于A区和B区。两个值落在C区,一个值落在D区。CGMS检测低血糖的敏感性和特异性分别为33%和96%。共记录了6666对传感器值,相关系数为0.84,变异系数为8.25%。

结论

CGMS在常规临床实践中可能有用,与间歇性血糖自我监测(SMBG)相比,它能提供更多关于血糖谱的信息。然而,CGMS不能替代血糖仪,因为它不满足为体外葡萄糖测量设定的传统性能目标,因此可能导致临床治疗决策错误。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验