连续血糖监测系统的临床性能评估:范围综述及报告建议。

Clinical Performance Evaluation of Continuous Glucose Monitoring Systems: A Scoping Review and Recommendations for Reporting.

机构信息

IFCC Scientific Division - Working Group on Continuous Glucose Monitoring.

Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.

出版信息

J Diabetes Sci Technol. 2023 Nov;17(6):1506-1526. doi: 10.1177/19322968231190941. Epub 2023 Aug 20.

Abstract

The use of different approaches for design and results presentation of studies for the clinical performance evaluation of continuous glucose monitoring (CGM) systems has long been recognized as a major challenge in comparing their results. However, a comprehensive characterization of the variability in study designs is currently unavailable. This article presents a scoping review of clinical CGM performance evaluations published between 2002 and 2022. Specifically, this review quantifies the prevalence of numerous options associated with various aspects of study design, including subject population, comparator (reference) method selection, testing procedures, and statistical accuracy evaluation. We found that there is a large variability in nearly all of those aspects and, in particular, in the characteristics of the comparator measurements. Furthermore, these characteristics as well as other crucial aspects of study design are often not reported in sufficient detail to allow an informed interpretation of study results. We therefore provide recommendations for reporting the general study design, CGM system use, comparator measurement approach, testing procedures, and data analysis/statistical performance evaluation. Additionally, this review aims to serve as a foundation for the development of a standardized CGM performance evaluation procedure, thereby supporting the goals and objectives of the Working Group on CGM established by the Scientific Division of the International Federation of Clinical Chemistry and Laboratory Medicine.

摘要

不同方法在连续血糖监测(CGM)系统临床性能评估研究的设计和结果呈现中一直被认为是比较其结果的主要挑战。然而,目前尚缺乏对研究设计中变异性的全面描述。本文对 2002 年至 2022 年间发表的临床 CGM 性能评估进行了范围综述。具体来说,本综述量化了与研究设计的各个方面相关的众多选项的流行程度,包括研究人群、比较(参考)方法选择、测试程序和统计准确性评估。我们发现,几乎所有这些方面都存在很大的变异性,特别是在比较测量的特征方面。此外,这些特征以及研究设计的其他关键方面通常没有详细报告,无法对研究结果进行明智的解释。因此,我们提供了报告一般研究设计、CGM 系统使用、比较测量方法、测试程序和数据分析/统计性能评估的建议。此外,本综述旨在为制定标准化 CGM 性能评估程序奠定基础,从而支持国际临床化学和实验室医学联合会科学分会设立的 CGM 工作组的目标和宗旨。

相似文献

4
9
Continuous glucose monitoring systems for type 1 diabetes mellitus.1型糖尿病的连续血糖监测系统
Cochrane Database Syst Rev. 2012 Jan 18;1(1):CD008101. doi: 10.1002/14651858.CD008101.pub2.

引用本文的文献

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验