Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure.

作者信息

Davies Justine I, Band Margaret M, Pringle Stuart, Ogston Simon, Struthers Allan D

机构信息

Department of Clinical Pharmacology, Department of Cardiology and Department of Epidemiology and Public Health, Ninewells Hospital and Medical School, Dundee DD1 9SY, Scotland, UK.

出版信息

J Hypertens. 2003 Mar;21(3):571-6. doi: 10.1097/00004872-200303000-00023.

Abstract

BACKGROUND

The Sphygmocor system purports to be able to assess ascending aortic blood pressure using a transfer function. It has been shown to be accurate when data obtained invasively are used, but has not been tested prospectively using data obtained non-invasively.

OBJECTIVE

To investigate the accuracy of this equipment when measurements are obtained non-invasively, as would normally be the case in the clinic setting.

DESIGN AND METHODS

The study was observational. Ascending aortic pressure measurements were taken simultaneously with radial artery pressure wave recordings for estimation of ascending aortic blood pressure, in 28 patients undergoing diagnostic cardiac catheterization.

RESULTS

The transfer function in the Sphygmocor system significantly underestimated invasively measured systolic blood pressure [mean -7.23 +/- 10.07 mmHg; 95% confidence interval (CI) -3.3 to -11.14 mmHg, P = 0.001] and significantly overestimated invasively measured diastolic blood pressure (mean 12.20 +/- 7.14 mmHg; 95% CI 9.43 to 14.97 mmHg, P <0.001). Oscillometrically measured brachial systolic blood pressure was not significantly different from that measured invasively in the ascending aorta (mean 3.36 +/- 10.47 mmHg; 95% CI -0.69 to 7.43 mmHg, P = 0.1), but oscillometric measurement of brachial diastolic blood pressure gave a significant overestimation of that measured invasively (mean 11.70 +/- 7.18 mmHg; 95% CI 8.91 to 14.49 mmHg, P <0.001). CONCLUSIONS The transfer function in the Sphygmocor system is no better at estimating ascending aortic blood pressure than are standard peripheral blood pressure measurements. It may be necessary to derive a new transfer system that is based on data that are acquired entirely non-invasively.