Saylor W L, Dillard M
Med Phys. 1976 Mar-Apr;3(2):117-9. doi: 10.1118/1.594277.
Measurements were made in water about the tandem and ovoid structures of the Fletcher-Suit applicator loaded with a 137Cs source. The dose rate was found to vary as much as 25% as a result of differential absorption in the ovoid. Empirically based trigonometric expressions are suggested which provide the absorption correction as a function of the two angles of a spherical coordinate system. Computer calculated dose rates, with and without the absorption corrections, are compared.
在装有¹³⁷Cs源的Fletcher-Suit施源器的串联和卵形结构周围的水中进行了测量。结果发现,由于卵形结构中的差异吸收,剂量率变化高达25%。提出了基于经验的三角表达式,该表达式根据球坐标系的两个角度提供吸收校正。比较了计算机计算的有无吸收校正的剂量率。