Blacha Jan
Katedra i Klinika Ortopedii, Traumatologii i Rehabilitacji, Akademia Medyczna w Lublinie.
Chir Narzadow Ruchu Ortop Pol. 2002;67(5):541-5.
The author presents a case of metal stem cone wear caused by a ceramic head. During revision hip arthroplasty procedure a new modular ceramic head was implanted in combination with a stable cemented TT-1 stem. 13 months after surgery the hip became symptomatic and 20 months after the procedure total loosening of the cup was observed. Wear of the stem cone was visible. During the next revision procedure all hip prosthesis elements were removed, soft tissue debridement was performed and a new cemented prosthesis was implanted. One year post-op the hip is pain free.
作者报告了一例由陶瓷股骨头导致金属柄锥磨损的病例。在髋关节翻修手术过程中,植入了一个新的模块化陶瓷股骨头,并结合使用了稳定的骨水泥固定TT-1柄。术后13个月,髋关节出现症状,术后20个月观察到髋臼杯完全松动。柄锥磨损明显。在下一次翻修手术中,移除了所有髋关节假体部件,进行了软组织清创,并植入了一个新的骨水泥固定假体。术后一年,髋关节无疼痛。