Department of Orthopaedics and Traumatology, Prince of Wales Hospital, The Chinese University of Hong Kong, Shatin, NT, Hong Kong, China.
J Arthroplasty. 2010 Jun;25(4):660.e5-7. doi: 10.1016/j.arth.2009.04.012. Epub 2009 Jun 2.
We present a case of early ceramic insert dislodgment in a young patient with revision uncemented ceramic-on-ceramic total hip arthroplasty. The presentation may be subtle and the dislodgment may go unnoticed. This incidence draws our attention to the need for careful intraoperative confirmation of the insert position and frequent serial radiographic monitoring. The case also draws speculation of possible design flaws in the locking mechanism predisposing for the insert dislodgment.