Montuschi P, Ragazzoni E, Valente S, Corbo G, Mondino C, Ciappi G, Barnes P J, Ciabattoni G
Department of Pharmacology, Catholic University of the Sacred Heart, L.go F. Vito, 1, Roma, Italy.
Inflamm Res. 2003 Feb;52(2):69-73. doi: 10.1007/s000110300003.
To qualitatively validate an enzyme immunoassay to measure leukotriene B4 in exhaled breath condensate. Exhaled breath condensate is a new non-invasive method to monitor airway inflammation.
Twenty-two subjects with different lung diseases attended the outpatient clinic on one occasion for exhaled breath condensate collection.
Samples were pooled together and purified by reverse-phase high-performance liquid chromatography. The fractions eluted were assayed for leukotriene B4 by enzyme immunoassay.
A single peak of leukotriene B4-like immunoreactivity co-eluting with leukotriene B4 standard (retention time: 24 min) was identified by enzyme immunoassay. Reverse phase-high performance liquid chromatography peak of leukotriene B4 was clearly separated from those of 6-trans-leukotriene B4 (retention time: 14 min) and leukotriene B5 (retention time: 18 min) for which the antiserum used in the enzyme immunoassay had the highest cross-reactivity. Leukotriene B4 recovery was 64%.
This study provides evidence for the presence of leukotriene B4 in the exhaled breath condensate and the specificity of the enzyme immunoassay used.
对一种用于检测呼出气冷凝液中白三烯B4的酶免疫测定法进行定性验证。呼出气冷凝液是一种监测气道炎症的新型非侵入性方法。
22名患有不同肺部疾病的受试者到门诊就诊一次,以收集呼出气冷凝液。
将样本汇集在一起,通过反相高效液相色谱法进行纯化。对洗脱的馏分采用酶免疫测定法检测白三烯B4。
通过酶免疫测定法鉴定出一个与白三烯B4标准品共洗脱(保留时间:24分钟)的白三烯B4样免疫反应单峰。白三烯B4的反相高效液相色谱峰与6-反式白三烯B4(保留时间:14分钟)和白三烯B5(保留时间:18分钟)的峰明显分离,而酶免疫测定中使用的抗血清对这两种物质具有最高的交叉反应性。白三烯B4的回收率为64%。
本研究为呼出气冷凝液中存在白三烯B4以及所用酶免疫测定法的特异性提供了证据。