Isocenter accuracy in frameless stereotactic radiotherapy using implanted fiducials.

作者信息

Kim Ki-Hwan, Cho Moon-June, Kim Jun-Sang, Kim Jae-Sung, Song Chang-Joon, Song Shi-Hun, Kim Seon-Hwan, Myers Lee, Kim Yong-Eun

机构信息

Department of Therapeutic Radiology, Chungnam National University, Taejon, South Korea.

出版信息

Int J Radiat Oncol Biol Phys. 2003 May 1;56(1):266-73. doi: 10.1016/s0360-3016(03)00088-9.

Abstract

PURPOSE

The stereotactic radiotherapy (SRT) system verifies isocenter accuracy in patient space. In this study, we evaluate isocenter accuracy in frameless SRT using implanted cranial gold markers.

MATERIALS AND METHODS

We performed frameless SRT on 43 intracranial tumor patients between August 1997 and December 2000. The treatment technique was determined by the tumor shape and volume, and by the location of critical organs. The coordinates of anterior-posterior and lateral port film were inputted to ISOLOC software, which calculated (1) the couch moves translation distance required to bring the target point to the isocenter, and (2) the intermarker distance comparisons between the CT study and the treatment machine films. We evaluated the isocenter deviation based on the error between orthogonal film target coordinates and isocenter coordinates.

RESULTS

The mean treatment isocenter deviations (x, y, z) were -0.03, 0.14, and -0.04 mm, respectively. The systematic component isocenter standard deviations were 0.28, 0.31, and 0.35 mm (1 SD), respectively, and the random component isocenter standard deviations were 0.53, 0.52, and 0.50 mm (1 SD), respectively.

CONCLUSIONS

The isocenter accuracy in the frameless SRT-implanted fiducial system is highly reliable and is comparable to that of other stereotactic radiosurgery systems.

摘要

目的

立体定向放射治疗(SRT)系统可在患者空间内验证等中心精度。在本研究中,我们使用植入的颅骨金标记物评估无框架SRT中的等中心精度。

材料与方法

我们于1997年8月至2000年12月期间对43例颅内肿瘤患者进行了无框架SRT。治疗技术根据肿瘤的形状和体积以及关键器官的位置来确定。将前后位和侧位端口片的坐标输入到ISOLOC软件中,该软件可计算:(1)将靶点移至等中心所需的治疗床平移距离;(2)CT研究与治疗机器胶片之间的标记物间距比较。我们根据正交胶片靶点坐标与等中心坐标之间的误差评估等中心偏差。

结果

平均治疗等中心偏差(x、y、z)分别为-0.03、0.14和-0.04mm。系统分量等中心标准差分别为0.28、0.31和0.35mm(1个标准差),随机分量等中心标准差分别为0.53、0.52和0.50mm(1个标准差)。

结论

无框架SRT植入基准系统中的等中心精度高度可靠,与其他立体定向放射外科系统相当。