Neu C P, Hull M L
Biomedical Engineering Graduate Group, University of California at Davis, Davis, CA, USA.
J Biomech Eng. 2003 Apr;125(2):180-8. doi: 10.1115/1.1560141.
Recent magnetic resonance imaging (MRI) techniques have shown potential for measuring non-uniform deformations throughout the volume (i.e. three-dimensional (3D) deformations) in small orthopedic tissues such as articular cartilage. However, to analyze cartilage deformation using MRI techniques, a system is required which can construct images from multiple acquisitions of MRI signals from the cartilage in both the underformed and deformed states. The objectives of the work reported in this article were to 1) design an apparatus that could apply highly repeatable cyclic compressive loads of 400 N and operate in the bore of an MRI scanner, 2) demonstrate that the apparatus and MRI scanner can be successfully integrated to observe 3D deformations in a phantom material, 3) use the apparatus to determine the load cycle necessary to achieve a steady-state deformation response in normal bovine articular cartilage samples using a flat-surfaced and nonporous indentor in unconfined compression. Composed of electronic and pneumatic components, the apparatus regulated pressure to a double-acting pneumatic cylinder so that (1) load-controlled compression cycles were applied to cartilage samples immersed in a saline bath, (2) loading and recovery periods within a cycle varied in time duration, and (3) load magnitude varied so that the stress applied to cartilage samples was within typical physiological ranges. In addition the apparatus allowed gating for MR image acquisition, and operation within the bore of an MRI scanner without creating image artifacts. The apparatus demonstrated high repeatability in load application with a standard deviation of 1.8% of the mean 400 N load applied. When the apparatus was integrated with an MRI scanner programmed with appropriate pulse sequences, images of a phantom material in both the underformed and deformed states were constructed by assembling data acquired through multiple signal acquisitions. Additionally, the number of cycles to reach a steady-state response in normal bovine articular cartilage was 49 for a total cycle duration of 5 seconds, but decreased to 33 and 27 for increasing total cycle durations of 10 and 15 seconds, respectively. Once the steady-state response was achieved, 95% of all displacements were within +/- 7.42 microns of the mean displacement, indicating that the displacement response to the cyclic loads was highly repeatable. With this performance, the MRI-loading apparatus system meets the requirements to create images of articular cartilage from which 3D deformation can be determined.
最近的磁共振成像(MRI)技术已显示出在测量诸如关节软骨等小型骨科组织的整个体积内(即三维(3D)变形)的非均匀变形方面的潜力。然而,要使用MRI技术分析软骨变形,需要一个能够根据在未变形和变形状态下从软骨采集的多个MRI信号构建图像的系统。本文所报道工作的目标是:1)设计一种能够施加400 N高度可重复循环压缩载荷并在MRI扫描仪的孔腔内运行的装置;2)证明该装置与MRI扫描仪能够成功集成,以观察模拟材料中的3D变形;3)使用该装置确定在无侧限压缩中使用平面无孔压头使正常牛关节软骨样本达到稳态变形响应所需的加载循环次数。该装置由电子和气动部件组成,对双作用气缸调节压力,以便:(1)对浸泡在盐水中的软骨样本施加负载控制的压缩循环;(2)一个循环内的加载和恢复时间持续时间不同;(3)负载大小变化,以使施加到软骨样本的应力在典型生理范围内。此外,该装置允许进行门控以采集MR图像,并能在MRI扫描仪的孔腔内运行而不产生图像伪影。该装置在施加负载方面表现出高度可重复性,所施加的平均400 N负载的标准偏差为1.8%。当该装置与编程有适当脉冲序列的MRI扫描仪集成时,通过组装通过多次信号采集获得的数据,构建了未变形和变形状态下模拟材料的图像。此外,正常牛关节软骨达到稳态响应的循环次数在总循环持续时间为5秒时为49次,但在总循环持续时间增加到10秒和15秒时分别降至33次和27次。一旦达到稳态响应,所有位移的95%在平均位移的±7.42微米范围内,表明对循环载荷的位移响应具有高度可重复性。凭借这种性能,MRI加载装置系统满足了创建可确定3D变形的关节软骨图像的要求。