Suppr超能文献

使用基于压电薄膜的记录设备检测睡眠中人类的磨牙症力量

Bruxism force detection by a piezoelectric film-based recording device in sleeping humans.

作者信息

Baba Kazuyoshi, Clark Glenn T, Watanabe Tatsutomi, Ohyama Takashi

机构信息

Removable Prosthodontics, Graduate School, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8549, Japan.

出版信息

J Orofac Pain. 2003 Winter;17(1):58-64.

Abstract

AIMS

To test the reliability and utility of a force-based bruxism detection system (Intra-Splint Force Detector [ISFD]) for multiple night recordings of forceful tooth-to-splint contacts in sleeping human subjects in their home environment.

METHODS

Bruxism-type forces, i.e., forceful tooth-to-splint contacts, during the night were recorded with this system in 12 subjects (6 bruxers and 6 controls) for 5 nights in their home environment; a laboratory-based nocturnal polysomnogram (NPSG) study was also performed on 1 of these subjects.

RESULTS

All 12 subjects were able to use the device without substantial difficulty on a nightly basis. The bruxer group exhibited bruxism events of significantly longer duration than the control group (27 seconds/hour versus 7.4 seconds/hour, P < .01). A NPSG study performed on 1 subject revealed that, when the masseter muscle electromyogram (EMG) was used as a "gold standard," the ISFD had a sensitivity of 0.89. The correlation coefficient between the duration of events detected by the ISFD and the EMG was also 0.89.

CONCLUSION

These results suggest that the ISFD is a system that can be used easily by the subjects and that has a reasonable reliability for bruxism detection as reflected in forceful tooth-to-splint contacts during sleep.

摘要

目的

测试一种基于力的磨牙症检测系统(牙内夹板力探测器[ISFD])在家庭环境中对睡眠中的人类受试者进行多晚有力牙齿与夹板接触记录的可靠性和实用性。

方法

使用该系统在12名受试者(6名磨牙症患者和6名对照者)的家庭环境中记录其夜间的磨牙症类型的力,即有力的牙齿与夹板接触,持续5个晚上;还对其中1名受试者进行了基于实验室的夜间多导睡眠图(NPSG)研究。

结果

所有12名受试者都能够在每晚基本没有困难地使用该设备。磨牙症患者组的磨牙症事件持续时间明显长于对照组(27秒/小时对7.4秒/小时,P <.01)。对1名受试者进行的NPSG研究表明,当将咬肌肌电图(EMG)用作“金标准”时,ISFD的灵敏度为0.89。ISFD检测到的事件持续时间与EMG之间的相关系数也为0.89。

结论

这些结果表明,ISFD是一种受试者可以轻松使用的系统,并且在睡眠期间通过有力的牙齿与夹板接触反映出,其对磨牙症检测具有合理的可靠性。