The effect of the physical characteristics of the carrier gas on the output of automatic plenum vaporizers was studied. The Vapor and Fluomatic halothane output at all dial settings was highest with oxygen as the carrier gas, as compared with helium and nitrous oxide and was a function of carrier gas viscosity at all dial settings. The output of the Fluotec Mark 2 at the 0.5 and 1% settings was highest with nitrous oxide as the carrier gas, but at 2, 3 and 4% settings it was highest with oxygen; at the 0.5% and 1% dial settings it was a function of carrier gas density, but at 2%, 3% and 4% it was a function of carrier gas viscosity.

摘要

研究了载气物理特性对自动充气式蒸发器输出的影响。与氦气和一氧化二氮相比,在所有刻度设置下,以氧气作为载气时,Vapor和Fluomatic氟烷的输出最高,并且在所有刻度设置下都是载气粘度的函数。在0.5%和1%刻度设置下,Fluotec Mark 2以一氧化二氮作为载气时输出最高,但在2%、3%和4%刻度设置下,以氧气作为载气时输出最高;在0.5%和1%刻度设置下,它是载气密度的函数,但在2%、3%和4%刻度设置下,它是载气粘度的函数。