Hawkins R C
Department of Pathology and Laboratory Medicine, Tan Tock Seng Hospital, 11 Jalan Tan Tock Seng, Singapore 308433.
Singapore Med J. 2003 Jan;44(1):8-11.
To assess four commercially available point-of-care HbA1c analytical systems with respect to (1) inaccuracy against the central laboratory HbA1c method and (2) imprecision against the HbA1c analytical goal of coefficient of variation < 3%.
Analytical inaccuracy was assessed by analysis of 110 patient samples on all five analytical platforms (Biorad Diastat, Drew DS5, Bayer DCA 2000, Nycomed Nycocard and Roche Tinaquant (used in central lab)). Analytical imprecision was assessed by analysis of two levels of patient sample four times daily for six days, as well as analysis of two levels of commercial control.
Deming linear regression for agreement: Diastat=0.98 x Tinaquant + 0.36; DS5=1.23 x Tinaquant - 0.65; DCA2000=0.95 x Tinaquant + 0.63; Nycocard=0.94 x Tinaquant + 0.92. Analytical coefficients of variation (CVs) at Tinaquant HbA1c levels of 6.2-10.8% were: Tinaquant 0.8-1.1%, Diastat 1.6-6.6%, DCA2000 2.6-7.2%, DS5 5.1-11.7%, Nycocard 8.5-15.3%. Two HbE samples gave elevated HbA1c results with the DS5 method.
The Diastat and DCA2000 systems gave the best performance with acceptable imprecision and good agreement with both the central lab and each other. The DS5 was less precise with a significant positive bias compared to the other methods and interference from HbE, while the Nycocard system showed the poorest precision in the evaluation. The Diastat and DCA2000 systems appear to be satisfactory analytical alternatives to both central laboratory (Tinaquant) testing and each other.
评估四种市售即时糖化血红蛋白(HbA1c)分析系统,涉及(1)相对于中心实验室HbA1c方法的不准确性,以及(2)相对于变异系数<3%的HbA1c分析目标的不精密度。
通过在所有五个分析平台(伯乐Diastat、德鲁DS5、拜耳DCA 2000、Nycomed Nycocard和罗氏Tinaquant(用于中心实验室))上分析110份患者样本评估分析不准确性。通过每天四次、持续六天分析两个水平的患者样本以及分析两个水平的商业对照来评估分析不精密度。
一致性的戴明线性回归:Diastat = 0.98×Tinaquant + 0.36;DS5 = 1.23×Tinaquant - 0.65;DCA2000 = 0.95×Tinaquant + 0.63;Nycocard = 0.94×Tinaquant + 0.92。Tinaquant HbA1c水平为6.2 - 10.8%时的分析变异系数(CVs)为:Tinaquant 0.8 - 1.1%,Diastat 1.6 - 6.6%,DCA2000 2.6 - 7.2%,DS5 5.1 - 11.7%,Nycocard 8.5 - 15.3%。两份HbE样本采用DS5方法时HbA1c结果升高。
Diastat和DCA2000系统表现最佳,不精密度可接受,与中心实验室以及彼此之间一致性良好。与其他方法相比,DS5精密度较低,存在显著正偏差且受HbE干扰,而Nycocard系统在评估中精密度最差。Diastat和DCA2000系统似乎是中心实验室(Tinaquant)检测以及彼此之间令人满意的分析替代方法。