We present NMR measurements of the diffusion of hyperpolarized 3He in the human lung performed at fields much lower than those of conventional MRI scanners. The measurements were made on standing subjects using homebuilt apparatus operating at 3mT. O(2)-limited transverse relaxation (T(2) up to 15-35s) could be measured in vivo. Accurate global diffusion measurements have been performed in vivo and in a plastic bag; the average apparent diffusion coefficient (ADC) in vivo was 14.2+/-0.6mm(2)/s, whereas the diffusion coefficient in the bag (3He diluted in N(2)) was 79.5+/-1mm(2)/s. 1D ADC mapping with high SNR ( approximately 200-300) demonstrates the real possibility of performing quality lung imaging at extremely low fields.
我们展示了在远低于传统MRI扫描仪场强的条件下,对超极化3He在人肺中扩散的核磁共振测量。这些测量是在站立的受试者身上使用自制设备在3mT场强下进行的。在体内可以测量到O(2)受限的横向弛豫(T(2)高达15 - 35秒)。已经在体内和塑料袋中进行了精确的整体扩散测量;体内的平均表观扩散系数(ADC)为14.2±0.6mm(2)/s,而在塑料袋中(3He在N(2)中稀释)的扩散系数为79.5±1mm(2)/s。具有高信噪比(约200 - 300)的一维ADC映射证明了在极低场强下进行高质量肺部成像的实际可能性。