Fini M, Savarino L, Nicoli Aldini N, Martini L, Giavaresi G, Rizzi G, Martini D, Ruggeri A, Giunti A, Giardino R
Experimental Surgery Department, Codivilla-Putti Research Institute, Rizzoli Orthopaedic Institute, via di Barbiano 1/10, 40136 Bologna, Italy.
Biomaterials. 2003 Aug;24(19):3183-92. doi: 10.1016/s0142-9612(03)00164-9.
The influence of fluorohydroxyapatite (FHA) coating and surface roughness of Ti6Al4V implants on bone response was investigated. Uncoated and FHA-coated screws with lower (LR and LR+FHA; Ra: 5.7+/-0.2 microm) and higher (HR and HR+FHA; Ra: 21.8+/-0.9 microm) surface roughness, were inserted into the diaphyses of 8 sheep tibiae. Twelve weeks after implantation, extraction torque and bone-to-implant contact were evaluated. The smoothest surfaces showed an improved extraction torque and significant differences were observed between LR and HR (-24.6%, p<0.0005), LR and HR+FHA (-30.7%, p<0.0005), LR+FHA and HR (-17.4%, p<0.005), and LR+FHA and HR+FHA (-24.0%, p<0.005). The bone-to-implant contact data paralleled the biomechanical data: the smoother the surface, the greater the bone-to-implant contact. Significant (p<0.0005) decreases in bone-to-implant contact were observed between LR+FHA and HR (-24.2%), and between LR+FHA and HR+FHA (-29.2%). The current findings suggest that LR surfaces significantly improve the osteointegration rate of implanted cortical screws independently of the FHA coating.
研究了氟羟基磷灰石(FHA)涂层和Ti6Al4V种植体表面粗糙度对骨反应的影响。将具有较低(LR和LR + FHA;Ra:5.7±0.2微米)和较高(HR和HR + FHA;Ra:21.8±0.9微米)表面粗糙度的未涂层和FHA涂层螺钉插入8只绵羊胫骨的骨干中。植入12周后,评估拔出扭矩和骨与种植体的接触情况。最光滑的表面显示拔出扭矩有所改善,并且在LR和HR之间(-24.6%,p<0.0005)、LR和HR + FHA之间(-30.7%,p<0.0005)、LR + FHA和HR之间(-17.4%,p<0.005)以及LR + FHA和HR + FHA之间(-24.0%,p<0.005)观察到显著差异。骨与种植体的接触数据与生物力学数据平行:表面越光滑,骨与种植体的接触越大。在LR + FHA和HR之间(-24.2%)以及LR + FHA和HR + FHA之间(-29.2%)观察到骨与种植体接触的显著(p<0.0005)下降。目前的研究结果表明,LR表面可显著提高植入皮质螺钉的骨整合率,且与FHA涂层无关。