Sermasi S, Marconi M
Department of Cardiology, Ospedale Infermi, Rimini, Italy.
Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 2):1903-7. doi: 10.1111/j.1540-8159.1992.tb02991.x.
Pacemaker mediated tachycardias (PMTs) are a well known complication of P synchronous pacing. Although the initiating mechanisms are several, all of them are associated with retrograde atrial activation, which is sensed by the atrial sensing channel, resulting in ventricular pacing. In 19 patients suffering from symptomatic AV conduction disturbances and normal sinus node function, a VDD pacing system connected to a single pass ventricular lead with dual chamber electrodes was implanted. The bipolar atrial electrode, floating in the right atrium, was used to detect endocardial atrial electrograms that were differentially processed within the pacemaker for optimal discrimination and filtering of undesirable signals. The widely programmable atrial sensitivity (amplitude and filtering) allowed stable P synchronized ventricular pacing in all patients, but in five of them, sustained PMTs not related to retrograde atrial activation was documented during the follow-up. The common mechanism for the onset and maintenance of these PMTs was traced to the abnormal sensing of the terminal forces of ventricular activation and/or of the T wave. The possibility of interferences between ventricular and atrial electrodes (crosstalk) was also considered. The reduction of atrial channel sensitivity represented in all cases the only effective procedure to prevent this type of PMT. In conclusion, the bet signal to noise ratio is an important endpoint to assure the proper function of a single lead VDD pacing system. Furthermore, using the differential amplifier built within the pacemaker, consideration should be given to the optimal mode of rejection of the terminal forces of the QRS and T wave.
起搏器介导的心动过速(PMT)是P波同步起搏的一种常见并发症。尽管引发机制有多种,但均与逆行心房激动有关,这种激动被心房感知通道感知,从而导致心室起搏。对19例有症状的房室传导障碍且窦房结功能正常的患者,植入了连接单根心室导线和双腔电极的VDD起搏系统。漂浮在右心房的双极心房电极用于检测心内膜心房电图,该电图在起搏器内进行差分处理,以实现对不良信号的最佳辨别和滤波。广泛可编程的心房灵敏度(幅度和滤波)使所有患者均能实现稳定的P波同步心室起搏,但在随访期间,其中5例记录到了与逆行心房激动无关的持续性PMT。这些PMT发作和维持的常见机制可追溯到心室激动终末力和/或T波的异常感知。还考虑了心室和心房电极之间干扰(串扰)的可能性。在所有病例中,降低心房通道灵敏度是预防此类PMT的唯一有效方法。总之,保证单根导线VDD起搏系统正常功能的一个重要终点是最佳信噪比。此外,使用起搏器内置的差分放大器时,应考虑对QRS波和T波终末力的最佳抑制模式。