Zhou J, Ens W, Standing K G, Verentchikov A
Department of Physics, University of Manitoba, Winnipeg, Canada.
Rapid Commun Mass Spectrom. 1992 Nov;6(11):671-8. doi: 10.1002/rcm.1290061109.
Measurements of kinetic energy distributions of molecular ions ejected into an extraction field by matrix-assisted laser desorption are reported. The measurements were made in a time-of-flight mass spectrometer with an electrostatic mirror by measuring the reflected signal as a function of the difference between the accelerating voltage and the voltage applied to the mirror. The molecular ions were found to have less kinetic energy than the extraction field alone would normally provide, i.e., we observed an energy deficit. Under conditions typical for a matrix-assisted laser desorption experiment, the deficit is about 24 eV for molecular ions of insulin. The size of the deficit increases with the intensity of the molecular ion signal, and the molecular weight of the protein; it is also larger for negative molecular ions than for positive molecular ions.
报道了通过基质辅助激光解吸喷射到提取场中的分子离子动能分布的测量结果。这些测量是在带有静电镜的飞行时间质谱仪中进行的,通过测量反射信号作为加速电压与施加到镜上的电压之差的函数。发现分子离子具有的动能比仅提取场通常提供的动能要少,即,我们观察到了能量亏缺。在基质辅助激光解吸实验的典型条件下,胰岛素分子离子的亏缺约为24电子伏特。亏缺的大小随着分子离子信号的强度和蛋白质的分子量增加;对于负分子离子,亏缺也比正分子离子更大。