Ens W, Mao Y, Mayer F, Standing K G
Department of Physics, University of Manitoba, Winnipeg, Canada.
Rapid Commun Mass Spectrom. 1991 Mar;5(3):117-23. doi: 10.1002/rcm.1290050306.
Some properties of matrix-assisted laser desorption have been studied using single-ion-counting methods and a time-to-digital converter. The methods allow examination of the process for irradiances near the reported threshold for observation with a transient recorder. All measurements were made using bovine insulin as a test compound. We present direct evidence that an irradiance threshold near 10(6) W cm-2 exists for ion production, and that the process is a collective effect, either involving a large number of molecular ions (approximately 10(4) in a successful event or none at all. Above the threshold, the yield is found to scale with a high power (4th to 6th) of the irradiance. Measurements of initial velocity distributions indicate an axial velocity spread corresponding to approximately 50 eV and a radial velocity spread corresponding to approximately 2.4 eV. Thus the ejection or extraction mechanism appears to be strongly asymmetric.