Ultrasound biomicroscopy and Scheimpflug photography of angle-supported phakic intraocular lens for high myopia.

作者信息

de Souza Renate Ferreira, Allemann Norma, Forseto Adriana, Barros Paulo S Moraes, Chamon Wallace, Nosé Walton

机构信息

Department of Ophthalmology, Federal University of São Paulo, Brazil.

出版信息

J Cataract Refract Surg. 2003 Jun;29(6):1159-66. doi: 10.1016/s0886-3350(02)02045-x.

Abstract

PURPOSE

To evaluate the equivalence of Scheimpflug photography (SP) and ultrasound biomicroscopy (UBM) in determining corneal epithelium-intraocular lens (IOL) and border IOL-iris distances.

SETTING

Universität Erlangen-Nuremberg, Erlangen, Germany.

METHODS

In 26 eyes of 17 patients who had a NuVita MA20 angle-supported anterior chamber intraocular lens (Chiron-Domilens), SP and UBM were used to evaluate the distance between the endothelium and the anterior IOL face in central and peripheral regions (12 o'clock and 6 o'clock positions) and between the border of the anterior IOL face and the iris. The Wilcoxon test was used for statistical analysis.

RESULTS

The mean central endothelium-anterior IOL face distance was 2.01 mm and 2.00 mm by SP and UBM, respectively. The mean peripheral endothelium-anterior IOL border distance was 1.28 mm and 1.58 mm, respectively, and the mean peripheral anterior IOL face-iris distance, 0.89 mm and 0.75 mm, respectively.

CONCLUSIONS

The difference between the 2 methods in the central endothelium-anterior IOL face distance was not significant (methods were equivalent), but the difference in the peripheral endothelium-anterior IOL face distance was. This may be the result of difficulty in obtaining the exact transition point between the IOL and the haptics by SP examination. The difference between the 2 methods in the IOL border-iris distance was also significant because of the irregularity of the iris surface; therefore, measurements were performed at different sites along this structure. The significant differences in the peripheral endothelium-IOL and IOL border-iris distances indicate that although both methods are useful, they are not equivalent.

摘要

目的

评估Scheimpflug摄影(SP)和超声生物显微镜检查(UBM)在测定角膜上皮-人工晶状体(IOL)以及IOL-虹膜边界距离方面的等效性。

地点

德国埃尔兰根-纽伦堡大学,埃尔兰根。

方法

对17例植入NuVita MA20角支撑型前房人工晶状体(Chiron-Domilens)的患者的26只眼,使用SP和UBM评估中央和周边区域(12点和6点位置)的内皮与IOL前表面之间的距离以及IOL前表面与虹膜之间的边界距离。采用Wilcoxon检验进行统计分析。

结果

通过SP和UBM测得的中央内皮-IOL前表面平均距离分别为2.01 mm和2.00 mm。周边内皮-IOL前边界平均距离分别为1.28 mm和1.58 mm,周边IOL前表面-虹膜平均距离分别为0.89 mm和0.75 mm。

结论

两种方法在中央内皮-IOL前表面距离上的差异不显著(方法等效),但在周边内皮-IOL前表面距离上存在差异。这可能是由于通过SP检查难以获得IOL与袢之间的确切过渡点所致。由于虹膜表面不规则,两种方法在IOL边界-虹膜距离上的差异也很显著;因此,沿着该结构在不同部位进行了测量。周边内皮-IOL和IOL边界-虹膜距离的显著差异表明,尽管两种方法都有用,但它们并不等效。